Hanna Instruments HI 3831F User Manual

Hi 3831f free chlorine test kit

Advertising
background image

SPECIFICATIONS

Instruction Manual

www.hannainst.com

Dear Customer,
Thank you for choosing a Hanna Product.
Please read the instructions carefully before using the
chemical test kit. It will provide you with the necessary
information for correct use of the kit.
Remove the chemical test kit from the packing material and
examine it carefully to make sure that no damage has
occurred during shipping. If there is any noticeable dam-
age, notify your Dealer or the nearest Hanna office
immediately.
Each kit is supplied with:

• 1 Color Comparator Cube;
• Reagent 1 (20 mL);
• Reagent 2 (15 mL).

Note: Any damaged or defective item must be returned in

its original packing materials.

ISTR3831FR1 10/01 PRINTED IN ITALY

HI 3831F

Free Chlorine

Test Kit

In pools and drinking water supplies, chlorination serves to
kill or deactivate disease-producing microorganisms. It can
also improve water quality by reacting with ammonia, iron,
sulfide and some organic substances. However, an excessive
concentration of chlorine in water can produce adverse
conditions, such as formation of carcinogenic chloroform or
other toxins. To maximize the purpose for chlorination and
minimize any adverse effects, it is essential to monitor the
chlorine levels closely.
The Hanna Chlorine Test Kit determines the Free chlorine
concentration in water via a color cube. This makes the test
kit practical for field use. No iodine or bromine can be
present for this test to work properly.

Range

0 to 2.5 mg/L (ppm) Chlorine

Smallest Increment

0.5 mg/L (ppm) Chlorine

Analysis Method

Colorimetric

Sample Size

5 mL

Number of Tests

50 (average)

Case Dimensions

220x145x55 mm (8.7x5.7x2.1")

Shipping Weight

176 g (6.6 oz.)

SPECIFICATIONS

SIGNIFICANCE AND USE

CHEMICAL REACTION

The addition of chlorine to water produces hydrochloric and
hypochlorous acids. The hypochlorous acid acts as the
disinfectant and bleaching agent. These are known as free
chlorine, which is measured by a colorimetric method. The
reaction is buffered at approximately 6.3 pH; in that
condition DPD (N,N-diethyl-p-phenylenediamine) is immedi-
ately oxidized by chlorine producing a reddish color. The
color intensity of the solution determines the free chlorine
concentration.

INSTRUCTIONS

READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
• Add 5 drops of Reagent 1 and 3 drops of Reagent 2 to

the color comparator cube.

• Fill the color comparator cube with water sample to the

5 mL mark.

• Replace the cap and mix by carefully swirling the cube

in tight circles and inverting it several times.

• Determine which color band best matches the solution in

the vessel and record the results in mg/L (ppm) free
chlorine.

REFERENCES

Standard Methods for the Examination of Water and Waste-
water, 18th Edition, 1992, pages 445-446.

The chemicals contained in this kit may be hazardous if
improperly handled. Read Health and Safety Data Sheet
before performing this test.

HEALTH AND SAFETY

5

3