HI 38082

Potassium

Test Kit for

Soil

www.hannainst.com

Dear Customer,
Thank you for choosing a Hanna Product.
Please read the instruction sheet carefully before using the
test kit. It will provide you with the necessary information
for correct use of the kit. If you need additional information,
do not hesitate to e-mail us at [email protected].
Remove the chemical test kit from the packing material and
examine it carefully to make sure that no damage has

occurred during shipping. If there is any noticeable dam-
age, notify your Dealer or the nearest Hanna office
immediately.
Each kit is supplied with:

• Formaldehyde Solution, 4 bottles with dropper (4 x 25 mL);
• HI 38082A-0 Alkaline Solution, 3 bottles (3 x 100 mL);
• HI 38082B-0 Potassium Reagent, packets (100 pcs);
• 1 long glass test tube (50 mL);
• 1 calibrated plastic vessel (50 mL);
• 1 long plastic pipette (1 mL);
• 1 plastic pipette (3 mL);
• 1 syringe (1 mL) with tip;
• 1 spoon;
• 1 graduated card;
• 1 point card.

Note: Any damaged or defective item must be returned in

its original packing materials.

ISTR38082 10/00 PRINTED IN ITALY

Potassium is present in tissues responsible for the growth of

plants (primary and secondary meristems). It plays an

important role in how much water is absorbed by the roots

and in the regulation of cellular activity. In addition,

potassium makes plants more resistant to diseases and

yields a positive effect on the color and fragrance in flowers.
The problem of lack of potassium is quite frequent in

calcareous soils.
Note: mg/L is equivalent to ppm (parts per million).

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
1- Obtain a Mehlich extract

from the soil sample by
performing the Mehlich
extraction.

2- Using the syringe, add

3 mL of soil extract to

the plastic vessel (a full

syringe contains 1 mL of

solution, hence repeat 3

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

The Hanna Test Kit determines Potassium in soil via a

turbidimetric method. Potassium is precipitated in a basic

environment with sodium tetraphenylborate. The developed

turbidity is proportional to potassium concentration.

times). Add deionized water up
to the 25 mL mark.

3- Add 18 drops of Formaldehyde

Solution and swirl to mix.

4- Using the short plastic pipette, add 3 mL of HI 38082A-0

Alkaline Solution and swirl to mix.

5- Add 1 packet of HI 38082B-0

reagent and mix by means of
the spoon.

6- Wait for 5 min-

utes to allow

reaction to com-

plete. If Potassium is present,

the solution will become tur-

bid.

7- Place the test tube on the

Point card

and look from the

top of the tube at the black
spot on the Point card. Use

the long pipette to fill the
tube with the reacted sample
until the black spot completely
disappears.

Range

0 to 50 mg/L (ppm) as K

50 to 250 mg/L (ppm) as K

Smallest Increment 5 mg/L [in the 0-50 mg/L range]

25 mg/L [in the 50-250 mg/L range]

Analysis Method

Turbidimetric

Sample Size

3 mL and 0.6 mL

Number of Tests

100

Case Dimensions

235x175x115 mm (9.2x6.9x4.5")

Shipping Weight

889.0 g (31.3 oz.)

8- Hold the tube close to the Graduated

card

as shown in the figure.

9- Read directly from the Graduated card

the concentration in mg/L (ppm) of po-

tassium that corresponds to the level of

the liquid in the test tube.

10- In case the black spot on the Point

card

disappears when the liquid

level is under the 50 ppm mark,
the potassium concentration is
higher than 50 ppm. Dilute the
original sample and perform the

test as follows.

11- Using the syringe, add to the plastic

vessel 0.6 mL of soil extract.

Note: To measure exactly 0.6 mL, take

the syringe and push the plunger
completely down into the syringe.
Insert tip into the extract and
pull the plunger out until the
lower edge of the seal is on the
0.0 mL mark of the syringe.

Insert the syringe into the vessel
and push the sample out until
the lower edge of the seal is on
the 0.6 mL mark.

x18

3 mL

0.6 mL

3 mL

25 mL