HI 38035

Total Hardness

and Calcium

Test Kit

Instruction Manual

ISTR38035 11/00 PRINTED IN ITALY

Water hardness has traditionally been defined as the
capacity of water to precipitate soap. The ionic species in
water causing the precipitation were later found to be
primarily calcium and magnesium. At the present time,
water hardness is a quantitative measure of these ions in
the water sample. Now it is also known that certain other
ion species, such as iron, zinc and manganese, contribute to

the overall hardness of water. The measure and subsequent
control of water hardness is essential to prevent scaling and
clogging in water pipes. Calcium concentration in water
depends on the source and on water treatment. Calcium
contributes to the total hardness of water.
The Hanna total hardness and calcium test kit makes
monitoring easy and quick. The compact size provides the
versatility to use the kit anywhere and the design makes

the kit easy to handle.
Note: 1 gpg (grains per gallon) CaCO

3

is equivalent to 17

ppm CaCO

3

(where

ppm - parts per million - is

equivalent to mg/L).

Range

0 to 20 gpg CaCO

3

as total hardness

0 to 20 gpg CaCO

3

as Ca

Smallest Increment

0.2 gpg CaCO

3

as total hardness

0.2 gpg CaCO

3

as Ca

Analysis Method

EDTA titration

Sample Size

25 mL

Number of Tests 100 (Total Hardness), 100 (Calcium)
Case Dimensions

370x270x80 mm (14.6x10.6x3.1")

Shipping Weight

960 g (33.9 oz.)

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

The total hardness level is determined by an EDTA (ethylene-

diamine-tetraacetic acid) titration. The solution is first adjusted
to a pH of 10 using a buffer solution. The indicator chelates
with metal ions such as magnesium or calcium to form a red
colored complex. As EDTA is added, metal ions complex with it.
After all the free metal ions have been complexed, an excess
EDTA removes the metal ions complexed with the indicator to
form a blue colored solution. This color change from red to

blue is the endpoint of the titration.

READ ALL THE INSTRUCTIONS BEFORE USING THE TEST KIT
Determination of Total Hardness
• Remove the cap from the plastic

vessel. Rinse the plastic vessel with
the water sample, fill to the 25 mL
mark and replace the cap.

• Add 5 drops of Buffer Solution through the cap port

and mix carefully swirling the vessel in tight circles.

• Add 1 drop of Calmagite Indicator through the cap port

and mix as described above. The solution becomes a red-
violet color.

• Take the syringe and push the

plunger completely down into the
syringe. Insert tip into the EDTA
solution gpg range and pull the
plunger out until the lower edge of

the seal is on the 0.0 mL mark of
the syringe.

• Place the syringe tip

into the cap port of
the plastic vessel and
slowly add the titra-
tion solution drop by
drop, swirling after each
drop.

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

Dear Customer,
Thank you for choosing a Hanna Product.
Please read the instruction sheet carefully before using the
test kit. It will provide you with the necessary information
for correct use of the kit. If you need additional information,
do not hesitate to e-mail us at [email protected].
Remove the chemical test kit from the packing material and
examine it carefully to make sure that no damage has
occurred during shipping. If there is any noticeable dam-

age, notify your Dealer or the nearest Hanna office
immediately.
Each kit is supplied with:

• Buffer Solution pH 10.2±0.2, 1 bottle with dropper

(30 mL);

• Calmagite Indicator, 1 bottle with dropper (10 mL);
• EDTA Solution 0-20 gpg range, 1 bottle (100 mL);
• HI 38035A-0 Calcium Reagent, 1 bottle with drop-

per (10 mL);

• HI 38035B-0 Calcium Reagent, 1 bottle with drop-

per (10 mL);

• HI 38035C-0 Calcium Reagent, 1 bottle with dropper

(15 mL);

• HI 38035D-0 Calcium Reagent, 1 bottle (100 mL);
• 1 calibrated plastic vessel (50 mL) with cap;
• 1 syringe (1 mL) with two tips.

Note: Any damaged or defective item must be returned in

its original packing materials.

25 mL

x 5

x 1

www.hannainst.com

• Continue adding the EDTA Solution until the solution in

the vessel becomes purple. Then mix for 15 seconds after
each additional drop until the solution in the vessel
turns blue.

• Read off the milliliters of titration solution used from the

syringe and multiply by 20 to obtain the hardness level
of your sample as gpg CaCO

3

.

mL of titrant x 20 = gpg CaCO

3

Determination of Calcium
• Remove the cap from the plastic

vessel. Rinse the plastic vessel with
the water sample, fill to the 25 mL
mark and replace the cap.

• Add 2 drops of HI 38035A-0 Calcium Reagent through

the cap port and mix carefully by swirling the vessel in
tight circles.

• Add 1 drop of HI 38035B-0 Calcium Reagent through

the cap port and mix as described above.

• Add 3 drops of HI 38035C-0 Calcium Reagent through

the cap port and mix as described above. The solution
will turn blue.

25 mL

x 1

x 3

x 2

Calcium concentration is determined by an EGTA titration.

Zincon indicator determines

a blue color in the solution. The

reaction endpoint is indicated by a change in color from
blue to orange.