Hanna Instruments HI 38020 User Manual

Instruction manual

Advertising
background image

Instruction Manual

HI 38020

Free & Total

Chlorine

Low, Medium and

High Range
Test Kit with

Checker Disc

www.hannainst.com

Dear Customer,
Thank you for choosing a Hanna Product.
Please read the instruction sheet carefully before using the

test kit. It will provide you with the necessary information

for correct use of the kit. If you need additional information,

do not hesitate to e-mail us at [email protected].
Remove the chemical test kit from the packing material and

examine it carefully to make sure that no damage has

occurred during shipping. If there is any noticeable dam-

age, notify your Dealer or the nearest Hanna office

immediately.
Each kit is supplied with:

• HI 93701-0 Free Chlorine Reagent, packets (100 pcs);
• HI 93711-0 Total Chlorine Reagent, packets (100 pcs);
• Demineralizer Bottle with filter cap for about 12 liters

of deionized water (depending on the hardness level

of water to be treated);

• 1 checker disc (including a mirror, a mirror holder ,

the 38020 and the 3875 discs);

• 2 glass vials with caps;
• 1 plastic pipette (3 mL).

Note: Any damaged or defective item must be returned in

its original packing materials.

ISTR38020 02/00 PRINTED IN ITALY

Chlorine is the most commonly used water disinfectant in
applications that vary from treatment of drinking water

and wastewater, pool and spa sanitization, to food pro-
cessing and sterilization. Chlorine present in water binds
with bacteria, leaving only a part of the original quantity
(free chlorine) to continue its disinfecting action. If the free
chlorine level is improper with respect to pH, water will
have an unpleasant taste and odor and the disinfecting
potential of the chlorine will be diminished.
Free chlorine reacts with ammonium ions and organic

compounds to form chlorine compounds resulting in dimin-
ished disinfecting capabilities compared with free chlorine.
Chlorine compounds together with chloramines form com-
bined chlorine. Combined chlorine and free chlorine together
result in total chlorine.
While free chlorine has a much higher disinfectant poten-
tial, combined chlorine has a much higher stability and
has a lesser volatility.
Note: mg/L is equivalent to ppm (parts per million).

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

The reaction between chlorine and the DPD reagent causes

a pink tint in the sample which is proportional to the
chlorine concentration.

READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
• To measure chlorine in the 0-0.7 mg/L range, verify

that the mirror, its holder and the 3875 disc are pre-

installed in the checker disc. If not, install them as

shown in the figure.

• Use the pipette to fill each glass

vial with 10 mL of sample (up to
the mark).

• Insert one of the vials into the left hand opening of the

checker disc. This is the blank.

• Remove the cap and fill

the Demineralizer Bottle
with tap water.

• Replace the cap and

shake gently for at
least 2 minutes. The
demineralized water

is now ready.

• Flip open the top of the Demineralizer Bottle cap. By

gently squeezing the bottle, add demineralized water
to the other vial up to the 20 mL mark. Replace the
cap and shake to mix.

10 mL

• Remove the cap add the specific test reagent:
Free Chlorine: 1 packet of

Total Chlorine: 1 packet of

DPD Free Chlorine reagent.

DPD Total Chlorine reagent.

• Replace the cap and mix.

• In the case of Total Chlo-

rine determination, wait
for 2 minutes to allow
reaction to occur. This is
the reacted sample.

• Remove the cap and insert

the reacted sample into the
right hand opening of the
checker disc.

• Hold the checker disc so that

a light source illuminates
the samples from the top.

• Keep the checker disc at a distance of 30-40 cm (12-

16") from the eyes to match the color. Rotate the disc
while looking at the color test windows and stop when
you find the color match. Read the value in the result
window and divide it by 5 to obtain mg/L (or ppm) of
Chlorine.

Range

0 to 0.7 mg/L (ppm) as Chlorine (LR)

0 to 3.5 mg/L (ppm) as Chlorine (MR)
0 to 10.0 mg/L (ppm) as Chlorine (HR)

Smallest Increment

LR: 0.02 mg/L

MR: 0.1 mg/L

HR: 0.5 mg/L

Analysis Method

Colorimetric

Sample Size

10 mL (LR), 5 mL (MR), 2.5 mL (HR)

Number of Tests

200

Case Dimensions

235x175x115 mm (9.2x6.9x4.5")

Shipping Weight

688 g (24.3 oz.)

Advertising