Chapter 3: Getting Started

Product Service 1-800-522-7658

8


Connecting to the Exhaust System (Optional)

ATTENTION: The weight of the exhaust ductwork system must
be supported independently of the enclosure superstructure or
damage may occur.

The exhaust system should be installed by a qualified HVAC
contractor
.

The exhaust connection on the TXE has been designed to accept 6" diameter
ductwork as shown in Figure 3-2. The exhaust connection ships uninstalled and
needs to be fastened with the screws provided. See Chapter 7 for ordering Canopy
Exhaust Connection for the TXE. Canopy Exhaust Connection aids in the removal
of chemicals or odors not captured by HEPA filters. Review Chapter 2 for exhaust
prerequisites and review Chapter 7 for ordering blower exhaust equipment.

Consult Labconco Customer Service should you require help sizing your blower
for the exhaust volume and system static pressure loss.

To ensure compatibility, the selected exhaust duct material
should match the enclosure, procedures and chemical
applications.



!

!