Chapter 6: Maintaining Your Filtered Enclosure

Product Service 1-800-522-7658

47

Determination of When and How to Replace Odor

Control Carbon Filters

The carbon filters MUST be replaced when any one of the following two
conditions are met:

1. The filtered enclosure outlet (exhaust) concentration approaches the inlet

concentration, indicating filter saturation.

2. The odor in the work area becomes intolerable or the concentration of the

chemical in the work area is greater than the TWA.

There are four means of determining when its time to change the carbon filters (not
shown in the order of preference).

Odor - A person’s sensitivity to odor, tolerance of odor and their comfort level
under odoriferous conditions vary with the individual. While odor is an indicator
that chemicals are passing through the carbon filter, several points need to be
understood:

Odor within the room is not necessarily an indication of saturation or
hazardous exposure concentrations.

Odor can be used as a prompt to sample the chemical concentration on the exit
side of the carbon filter.

Organic chemicals approved for use in the filtered enclosure have odors that
are detectable before reaching the time weighted exposure limits.

Detection Tubes - Color change indicators can be used to measure the
concentration of the chemical at the exit side of the carbon filter or in the outlet
exhaust. A kit including syringe pump and flexible tubing can be purchased as an
accessory from Labconco (Catalog # 6924900). Labconco Customer Service
Representatives are supplied with detector tube catalog numbers, as well as
telephone numbers to direct you to where to purchase these items.

For Organic, Formaldehyde and Ammonia, chemical specific detector tubes should
be purchased when installing fresh filters. Each kit contains instructions on how
many strokes of the syringe are required to obtain the stated sensitivity. The
sampling syringe is connected to the filtered enclosure exhaust. Connect the
syringe to the detector tube while the system is running and pull the air through the
tube with the syringe. Each stroke of the syringe represents a 100-ml sample and
corresponds to the number of strokes necessary to give the indicated color changes.
Due to the wide variety of organics and varying TWA’s, it is recommended that
specific detector tubes be purchased directly from Sensidyne, Draeger or your
laboratory supply dealer. Alternate detector pumps can also be purchased from
your laboratory supply dealers.

The vast majority of detector tubes available start measuring at the TWA. When a
user observes a color change in the tube, they should replace the filter
immediately. If no detector tube for your specific chemical is available, other
means of detection must be used.

Advertising
This manual is related to the following products:

XPert Filtered Balance Stations 3962621, XPert Filtered Balance Stations 3962620, XPert Filtered Balance Stations 3962602, XPert Filtered Balance Stations 3962600, XPert Filtered Balance Stations 3962522, XPert Filtered Balance Stations 3962521, XPert Filtered Balance Stations 3962520, XPert Filtered Balance Stations 3962502, XPert Filtered Balance Stations 3962501, XPert Filtered Balance Stations 3962500, XPert Filtered Balance Stations 3940422, XPert Filtered Balance Stations 3940421, XPert Filtered Balance Stations 3940420, XPert Filtered Balance Stations 3940402, XPert Filtered Balance Stations 3940401, XPert Filtered Balance Stations 3940400, XPert Filtered Balance Stations 3940321, XPert Filtered Balance Stations 3940320, XPert Filtered Balance Stations 3940302, XPert Filtered Balance Stations 3940301, XPert Filtered Balance Stations 3940300, XPert Filtered Balance Stations 3940222, XPert Filtered Balance Stations 3940221, XPert Filtered Balance Stations 3940220, XPert Filtered Balance Stations 3940202, XPert Filtered Balance Stations 3940201, XPert Filtered Balance Stations 3940200, XPert Filtered Balance Systems 3971402, XPert Filtered Balance Systems 3971401, XPert Filtered Balance Systems 3971400, XPert Filtered Balance Systems 3971302, XPert Filtered Balance Systems 3971301, XPert Filtered Balance Systems 3971300, XPert Filtered Balance Systems 3971202, XPert Filtered Balance Systems 3971201, XPert Filtered Balance Systems 3971200, XPert Filtered Balance Systems 3961622, XPert Filtered Balance Systems 3961621, XPert Filtered Balance Systems 3961620, XPert Filtered Balance Systems 3961602, XPert Filtered Balance Systems 3961601, XPert Filtered Balance Systems 3961600, XPert Filtered Balance Systems 3961522, XPert Filtered Balance Systems 3961521, XPert Filtered Balance Systems 3961520, XPert Filtered Balance Systems 3961502, XPert Filtered Balance Systems 3961501, XPert Filtered Balance Systems 3961500, XPert Filtered Balance Systems 3950422, XPert Filtered Balance Systems 3950421, XPert Filtered Balance Systems 3950420, XPert Filtered Balance Systems 3950402, XPert Filtered Balance Systems 3950401, XPert Filtered Balance Systems 3950400, XPert Filtered Balance Systems 3950322, XPert Filtered Balance Systems 3950321, XPert Filtered Balance Systems 3950320, XPert Filtered Balance Systems 3950302, XPert Filtered Balance Systems 3950301, XPert Filtered Balance Systems 3950300, XPert Filtered Balance Systems 3950222, XPert Filtered Balance Systems 3950221, XPert Filtered Balance Systems 3950220, XPert Filtered Balance Systems 3950202, XPert Filtered Balance Systems 3950201, XPert Filtered Balance Systems 3950200