Working in the biosafety cabinet – Labconco Biological Safety Cabinets 30268 Series User Manual

Page 42

Advertising
background image

Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

36

Original instructions

Working In the Biosafety Cabinet

Note: A more thorough review of using the BSC can be found in: Biosafety
in Microbiological and Biomedical Laboratories (BMBL)
, Published
by the Centers for Disease Control and Prevention
(

www.cdc.gov/biosafety/publications

).

Planning

• Thoroughly understand procedures and equipment required before

beginning work.

• Arrange for minimal disruptions, such as room traffic or entry into the

room while the cabinet is in use.

Start-up

• Turn off UV light if included.

• Slowly raise the sash until the bottom of the sash aligns with the sash

indicator decal located on the left side of the work area.

• Turn on the fluorescent light and cabinet blower if the Smart-Start

features have not been activated.

• Check the air grilles for obstructions.

• Allow the cabinet to operate until the display screen is shown.

Warming

up

Screen Display

Screen

• Wash hands and arms thoroughly with germicidal soap.

• Wear appropriate personnel protective equipment (PPE).

Wipe-Down

• Raise the sash to its full open position (approximately 21.75 inches or

552 mm). Mute the alarm by depressing the "OK/Mute" switch.

• Wipe down the interior surfaces of the cabinet with 70% ethanol, or a

suitable disinfectant, and allow to dry.

Advertising