Refer to the following table if the Biosafety Cabinet fails to operate properly.
If the suggested corrective actions do not solve the problem, contact
Labconco for additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Cabinet blower and
lights won’t turn on

Unit not plugged into
outlet

Plug the Biosafety Cabinet into
appropriate electrical service.

Check connection to control box on top
of Cabinet.

System Reset Switch
is Off

Turn on the System Reset Switch.

Circuit

breaker(s)

tripped

Reset circuit breakers.

Keypad

disconnected

or defective

Run keypad diagnostics and check
connections.

Blower won’t turn
on but lights work

Sash closed

Raise sash.

Blower wiring is
disconnected

Inspect blower wiring.

Blower

motor

is

defective

Replace blower motor.

Keypad

disconnected

or defective

Run keypad diagnostics and check
connections.