Chapter 4: Performance Features and Safety Precautions

Product Service 1-800-522-7658

24

Original instructions

Proper operation of the cabinet depends largely upon its location and the
operator’s work habits. Consult the Installation and Normal Operation
sections of this manual for further details.
Avoid direct exposure of plastic or coated materials to ultraviolet (UV)
radiation. Never bypass the UV safety interlock that only allows the UV light
to work when the sash is closed. When surface disinfecting the biosafety
cabinet:

• Avoid splashing the disinfecting solution on skin or clothing.

• Ensure adequate ventilation.

• Carefully follow the disinfectant’s safety instructions.

• Always dispose of disinfecting solutions in accordance with local and

national laws.

• DO NOT allow disinfectants with high concentrations of free chlorine

to contact the stainless steel components of the biosafety cabinet for a
long period of time. Free chlorine will corrode stainless steel after
extended contact.

Biosafety cabinets should be decontaminated for any of the following
reasons:

• Before maintenance work requiring entry into contaminated areas.

• Before HEPA filter changes.

• Before performing certification tests requiring entry into

contaminated areas.

• Before relocating the cabinet.

• Before changing research programs.

• After the gross spill of biohazardous material or toxic chemicals.