Chapter 3: Getting Started

Figure 3-2


All wiring for the fume hood SHOULD be
performed by a licensed electrician and conform
to all local codes. In most cases, the hood will
require the use of shielded conduit to protect the
wiring into the hood. The grounding connection
shall not be made to the terminal box cover.

The fluorescent light has been mounted outside the top liner panel
and is sealed from vapors inside the hood structure. To change the
fluorescent light bulbs in your hood, you must first remove the
front panel from the hood. Next remove the knock out plugs
holding the light fixture in place. Lift fixture up and replace any
defective bulbs. Reverse order to reassemble.

Connecting the Service Lines to
the Protector XStream Fume Hood

The hoods with service fixtures have been plumbed from the valve
to the hose connector or gooseneck for your installation
convenience. A qualified installer shall provide the tubing.
Tubing can enter the hood from above, through the back, or
through the work surface to make these connections to the service
fixtures.

Internal Junction
Box

Primary I.D. Plate
Located behind
Front Panel

Front Panel

Secondary I.D.
Label