Chapter 7: Troubleshooting

Product Service 1-800-522-7658

52

PROBLEM

CAUSE

CORRECTIVE ACTION

Cabinet blower
turns on but lights
don’t work

Lamps not installed
correctly

Inspect lamp installation.

Lamp is defective

Replace lamp.

Keypad

disconnected

or defective

Run keypad diagnostics and check
connections.

Lamp wiring is
disconnected

Inspect lamp wiring.

Defective

lamp

ballasts

Replace lamp ballasts.

Keypad

disconnected

or defective

Run keypad diagnostics and check
connections.

HEPA filter loading

The gauge reading will steadily
decrease as the cabinet is used.

Blockage of the
return air slots or
grille

Check all return air slots and grilles to
ensure that they are not blocked or
restricted.

Blockage of the
exhaust outlet

Ensure that the exhaust outlet is not
blocked or restricted.

Airflow Alert goes
off and/or there is a
slight decrease in
filter life remaining
gauge

Blockage or
restriction under the
work surface

Ensure that the towel catch and plenum
beneath the work surface are
unobstructed.