Labconco Instruction Sheet 5244201, Rev. -, ECO E275 3 of 6

Life Span for Oxygen Removal Column & Moisture Trap Removal

Note: Precise Glove Box Volume is 13 cu. ft. or 368 liters; 3000ppm or 0.3% of 368 liters = 1.1 liters of oxygen initially used

Glove Box

Initial Oxygen

Catalog# 5244200

Total Oxygen

Total Oxygen

Number

Main Chamber Pressure

Oxygen Leak Rate

Capacity Used

Oxygen Column Life

Life Capacity

Volume Capacity of Columns

(inches of water)

(ppm/min)

(ppm)

(24 hour days)

(ppm)

(liters)

-1

0.52

3000

21

19022

7

1

0

0.22

3000

51

19022

7

1

1

0.15

3000

74

19022

7

1

Note: All data with glove port covers installed

Glove Box

Catalog# 5244200

Total Moisture

Total Moisture

Number

Main Chamber Pressure Moisture Leak Rate

Initial Moisture

Moisture Column Life

Life Capacity

Weight Capacity of Columns

(inches of water)

(ppm/min)

(ppm)

(24 hour days)

(ppm)

(grams of moisture)

-1

2.9

10000

3318

13864241

100

1

0

1.7

10000

5659

13864241

100

1

1

1.6

10000

6013

13864241

100

1

Note: 1ppm of moisture = 7.21ug of moisture inside the 368 liter glove box volume

Note: All data with glove port covers installed


Moisture/Oxygen Leak Rate @ -1, 0, +1 Box Pressure

0

50

100

150

200

250

300

350

400

0

10

20

30

40

50

60

70

80

Time (Minutes)

p

p

m

Mo

ist

u

re/O

xyg

e

n

Moisture -1

2.9 ppm/min

Moisture 0

1.7 ppm/min

Moisture +1

1.6 ppm/min

Oxygen -1

0.52ppm/min

Oxygen 0

0.22 ppm/min

Oxygen +1

0.15 ppm/min