Labconco Guardian 500 Airflow Monitor User Manual

Page 4

Advertising
background image

Labconco Instruction Sheet 3909501, Rev. B, ECO F597 4 of 6

Calibration Procedure

Each alarm module and enclosure/fume hood is unique and needs to be individually calibrated in
the field. The procedure for the adjustment is as follows:

1. Double check the installation to make sure that monitor and power supply are properly

installed.

2. Allow 10 minutes for the monitor to warm up once power has been connected.

3. Determine the low flow set point for your monitor. This is the value where the monitor

will first indicate a low flow condition. The red light will be on for this value. Refer to
your industrial hygiene officer for the proper low flow set point or consult the table
below.

4. Adjust your enclosure/fume hood airflow to the low flow set point as previously

determined. The exhaust flow can be lowered by adjusting the speed control on the
FilterMate or by using an adjustable damper on the exhaust blower. Typical alarm
conditions are set at face velocities of 10 to 20 feet per minute below the normal
operating conditions due to supply air and exhaust air fluctuations, as well as room air
cross drafts. See note 8 if the low airflow volume or sash opening cannot be adjusted.

5. Using a properly calibrated thermoanemometer, determine the velocity through the face

of the enclosure by taking a detailed velocity traverse. Divide the face area into equal
increments. One reading per square foot of face area is normally recommended for an
accurate traverse. Compute the average velocity for this area.

6. If the red light alarm is on, slowly turn the adjustment screw counterclockwise until the

green light is activated. If the green light is on, slowly turn the adjustment screw
clockwise until the red light comes on. Slowly turn the adjustment screw back until the
red light is activated. It is important that these adjustments be done in small increments,
at intervals about 30 seconds apart to allow for delayed reaction of the alarm itself. The
alarm low flow set point should now be set and the red light activated.

7. Readjust the enclosure airflow to its normal operating levels. The green light should now

be activated. Calibration is now complete.

Enclosure Operating In Flow

Speed

Alarm Condition Set Point

Speed

100 fpm

80 - 90 fpm

80 fpm

60 - 70 fpm

60 fpm

40 - 50 fpm

8. Note: If the low airflow volume or sash opening cannot be adjusted, then a 1/4 to 1/3 of a

turn counterclockwise can be adjusted to set the airflow alarm condition at 10-25% below
normal operating levels.

Guardian 500 Alarm Activation


The audio and visual alarm will activate approximately 10-30 seconds after an alarm condition is
detected. To temporarily mute the audible alarm, press and release the test/reset button.

NOTE: After an alarm condition has been detected, the red light will stay on. The audible
alarm will remain muted until airflow returns to normal levels.

Advertising