Troubleshooting – Labconco Guardian Jr. Airflow Monitors 4865001 User Manual

Page 24

Advertising
background image

TROUBLESHOOTING

21

Symptom

Suggested

Recommendations

No Lights

Power supply not plugged into proper voltage; plug in

power supply. Monitor rear connection is
disconnected; verify that the connector is correctly
installed.

No Audible Alarm

Alarm has been silenced using test/reset button.

Wrong Alarm Set Point

Potentiometer was not properly adjusted. Repeat

calibration steps outlined in this manual.

Continuous Alarm

Blower speed has changed. Performance traverse to

verify that the calibration has not changed. Check
blower speed, adjust if required. Check calibration
using traverse technique, recalibrate monitor as
instructed if required.

Audible Disable Will
Not Stay On

An alarm condition must be continuously present

before the audible alarm can be silenced. If flow
conditions fluctuate near the alarm set point, the alarm
will automatically reset itself. Action should be taken
to bring the fume hood airflow into proper operating
parameters.

Constant Audible Alarm

Check airflow and calibration.


If problems persist with the Guardian Jr. Airflow Monitor, contact Labconco Product Service
directly at 1-800-522-7658, for assistance.