Please read these Directions for Use carefully to become thoroughly familiar

with proper operation and maintenance of this bulb suction before using it.

NeoNatalie Suction

Newborn - Reusable

Bulb Suction / Bulb Syringe

Directions for Use

Cat.No. 986000

QTY

1 each

Manufactured in China for:

Laerdal Medical AS, Tanke Svilandsgate 30
P.O. Box 377, 4002 Stavanger Norway

Tel. (+47) 51511700

Fax (+47) 51523557

E-mail: [email protected]

© 2010, Laerdal Medical AS. All rights reserved.

Distributed in the USA by:
LAERDAL MEDICAL CORPORATION
167 Myers Corners Road, P.O. Box 1840
Wappingers Falls, New York 12590-8840

Tel. (800) 431-1055, +1 (845) 297-7770
Fax (800) 227-1143, +1 (845) 298-4545
E-mail: [email protected]

20-02076 Rev E

Cleaning After Use

1. Open the Head part and empty remaining suctioned matter.

2. Wash and scrub all parts in soapy water. Flush water through the Nozzle

and Head. Ensure that all visible contamination is removed.

3. Disinfect by one of the following methods, listed in descending order of

efficacy:

Steam autoclaving: 136 °C (277 °F) and 2.0 kg/cm

2

for 10 - 20 minutes

Chemical disinfection: Activated glutaraldehyde solution in accordance

with manufacturer's recommended concentration and procedure. Rinse

in clean water by totally immersing the device for 1 minute.

Boiling in water: Minimum 10 minutes.

4. Dry completely and keep clean until next use.

Technical Specifications

Nozzle dimensions at tip: Inner diameter (ID): 3.0 mm

Outer diameter (OD): 4.5 mm

Suction strength(typical): 100 mmHg [136 cmH

2

O]

Operating temperature: 0 °C (32 °F) to 50 °C (122 °F)

Storage temperature:

-20 °C (-4 °F) to 60 °C (140 °F).

Material: Silicone