Connecting the Blood Pressure Cuff

The Patient Simulator is delivered with a specially adjusted blood
pressure cuff. Connect the tube to the white BP connector at the
side of the Patient Simulator before use.

Calibrating the Blood Pressure Cuff
Using LLEAP

1 Select <Tools> <Maintenance> and choose <Calibrate BP...>

2 Follow the onscreen wizard instructions to perform the

calibration.

Connecting the SpO

2

Probe

The SimMan 3G SpO

2

probe is

made up of a light diode and
light sensor. When the beam
between the diode and sensor
is broken, the Patient Monitor
Application registers that the
SpO

2

probe is connected.

1 Connect the probe’s USB

plug to the Patient Monitor
PC.

2 The probe can be placed on any suitable area on the Patient

Simulator, Ensure that the probe is always firmly fixed in position.