MAINTENANCE

Attaching the Right Leg

Please take the same precautions as when attaching the Left Leg.

1 Align the Right Leg bolt and cables with the pelvis socket. Feed

the leg bolt and cabling through the pelvis socket.

2 Carefully push the leg in towards the pelvis to form a snug fit.

3 Secure the cables and tubes in the connector. Screw the

connector in place on the leg bolt.

4 Connect the corresponding tubes and cables as shown in the

table below:

Right Leg to Pelvis – Cable and Tube Descriptions

Name/Label

Tube/Cable Color

Connector Description

Blood Right leg

Transparent, red
tube

Black twist and lock
connector

Right leg

Black harness cable

Black rectangular
connector, 4 lead

Air from leg

Colourless
transparent tube

White twist and lock
connector

Fluid from leg

Tube, blue

White twist and lock
connector

Attaching the Left Arm

Open the torso as described in steps 1 - 4, Opening the Torso.
Follow the procedures listed below in reverse to detach the arms.

1 Align the Left Arm axle with the shoulder socket.

2 Ensure that the shoulder screw is loose enough to allow the arm

axle to slide easily into place.

3 Feed the cables from the arm axle through the shoulder socket.

4 Carefully push the arm axle into the shoulder bracket, so that

the axle is flush with the inside of the bracket.

5 Tighten the shoulder screw with the Allen wrench.

LA PULSES

BP

LA

CONV

P

B

L

BP

PULSES

L

L

LA

B

6 Connect the arm cables to the corresponding connection points

in the torso.

Left Arm to Torso – Cable and Tube Descriptions

Name/Label

Tube/Cable Color

Connector Description

LA Pulses

Grey cable

Black rectangular
connector, 6 lead

BP

Grey cable

Black rectangular
connector, 2 lead