OPERATING INSTRUCTIONS CONTINUED

Current may be set below 2.0 amperes but actual discharge current

and/or meter reading may be inaccurate. Actual discharge amperes

may be read using the ampere test points (#10) with a digital multi-

meter set to the DC milli-volt scale.

Subsequent amperes will default to the previous test setting. If

different amperes are required from the previous test, the discharge

current must be set to the new current.

6. Press ‘Start’ (#2)

‘Testing’ LED illuminates

Cooling fan starts

Ammeter reading ramps to approximately the preset level then settles

to the steady preset level within +/- .1 amp.

Voltmeter reading drops to battery loaded voltage level.

Timer begins counting time.

NOTE: CURRENT SETTING CANNOT BE CHANGED WHILE TEST IS PROCEEDING.

UNIT MUST BE SHUT-OFF AND RESET WITH SWITCH (#1)

7. Test proceeds with battery voltage steadily decreasing.

8. When battery voltage lowers to the preset level (Step 3) the test is

complete.

The level switch may be set to the alternate level during the test.

(E.g. from 20 to 18 V, or 18 to 20 V)

Test complete (#9) LED illuminates

All readings freeze at their final readings.