Calibration, Indicator description – LW Measurements LSS 400 Large Shipping Scale User Manual

Page 3

Advertising
background image

CALIBRATION

When to calibrate - calibration is almost never required.

If the scale is inaccurate, calibration may be desired when the scale is first set

up for use, or if the scale is moved to a different altitude or gravitation. This is

necessary because the weight of a mass in one location is not necessarily the

same in another location. Also, with time and use, mechanical deviations can

occur.

How to calibrate

Note: calibration value must be more 2/3 of the capacity value that you

have set.

1. Press and keep 【ZERO】 key then press ON/OFF key to turn on the scale,

the display will show XXX.XX kg

2. Press 【UNIT】 key to select setting unit between kg and lb , press 【HOLD】

key to move flash digit to right, press 【ZERO】 key to increase flash digit..

Press 【TARE】 key, the display will show CAL then AD value. Press

【UNIT】 key, the display will show flashed calibration weight. Place the

shown weight then press 【UNIT】 key, the display will show ------- then AD

value. Take off calibration weight then turn off the scale. Calibration is

finished.

If calibration is repeatedly incorrect then: 1) Replace the batteries, 2) Try a more

stable surface or vibration and interference free location.










5

INDICATOR DESCRIPTION

BACK

AC adaptor

Socket

ON/OFF key

Turn the scale

power on/off.

HOLD key

The item’s weight reading will remain

on the display

UNIT key

Change weighing units, e.g. Kg or lb,.

TARE key

Tare Range (100% of full capacity)

ZERO key

Set display to zero or
Zero Range (4% to full capacity)

DISPLAY

Low battery indicator Weighing unit

Negative value indicator






ZERO indicator Hold indicator
Indicates when the reading is 0


2

ON/OFF

HOLD

UNIT

ZERO

TARE