Function settings, Calibration – LW Measurements Remo-R Indicator User Manual

Page 3

Advertising
background image

FUNCTION SETTINGS

1. Press and hold [HOLD] key then press [ON/OFF] key to turn on the scale, the

display will show A-ON or A-OFF, press [ZERO] key to select on or off (ON

means auto power off function activate, OFF means auto power off function

inactivate).

2. Press [UNIT] key, the display will show L-ON or L-OFF, press [ZERO] key to

select on or off (ON means auto backlight function activate, OFF means auto

backlight function inactivate).

3. Press [UNIT] key, the display will show P-XX. Press ZERO key to select OFF,

key, con and Stb. (OFF means RS232 function inactivates, KEY means that

output the data by pressing [PRINT] key. Con means that output the data

continually. Stb means that output the data when reading is stable)

4. Press [UNIT] key, the display will show J-ON or J-OFF, press ZERO key to

select ON or OFF. (ON means showing zero when turn on the power, OFF

means showing weight placed when turn on the power.)

5. Press [UNIT] key, the display will show F-Nor or F-PEA or F-Ani. Press ZERO

key to select Nor or PEA or Ani. (Nor means normal weighting function activate.

PEA means peak hold weighing function activate. Ani means animal weighting

function activate).

6. Press [UNIT] key, the display will show N-dsp or N-Ind. Press ZERO key to

select dsp or Ind. (dsp means it can display synchronously as a displayer for

another indicator through the RS232 socket. Ind means it is only a indicator for

the scale)

7. Press [UNIT] key, the display will show Kd 50 or Kd 100 or Kd 200 or Kd400.

(

press [ZERO] key to select animal scale anti-shock range among 50, 100,

200, 400)


5

Calibration :

1. Press and hold [ZERO] key then press [ON/OFF] key, the display will show

CAPU=, press [UNIT] key to select full capacity unit kg or lb(unit LB cannot be

selected if the LB is disabled in FUNCTION SETTING), press [HOLD] key to

confirm the selection.

2. Press [HOLD] key again to enter next menu, the display will show CAP, press

[UNIT] key to set full capacity, press [UNIT] key to move the flash digit to right,

press [ZERO] key to increase the flash digit, press [HOLD] key to confirm the

selection.

3. Press [HOLD] key again to enter next menu, the display will show CALU=,

press [UNIT] key to select calibration unit kg or lb(unit LB cannot be selected if

the LB is disabled in FUNCTION SETTING), press [HOLD] key to confirm the

selection.

4. Press [HOLD] key again to enter next menu, the display will show CAL, press

[UNIT] key to set calibration weight ,press [UNIT] key to move the flash digit to

right, press [ZERO] key to increase the flash digit, press [HOLD] key to confirm

the selection (the calibration weight should be more than 2/3 of full capacity).

5. Press [HOLD] key again to enter next menu, the display will show DIV, press

[UNIT] key to select division, press [HOLD] key to confirm the selection.

6. Press [HOLD] key again to enter next menu, the display will show Adb, press

[UNIT] key to select (19.20), press [HOLD] key to confirm the selection. Now

start the calibration as following steps:

Press [PRINT] key, the display will show CAL, then AD value, press [UNIT] key

when the stable indicator lights, the display will show flash calibration weight,

place the shown weight on the platform, press [UNIT] key when the stable

indicator lights, the display will show -----, and AD value. Turn off the scale now,

calibration is finished.

Turn on the scale, place tested weight to check whether it is correct, if not,

please repeat above segments.