7 Malfunctions and error correction

74

Version 01/2011, Rev. 1.5 of 16/12/2013 • sb

Translation of the original operating manual

Pos: 167 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 20 Kei n au sreic hend es Vak uum @ 6\ mod _12 991 375 151 07_ 68.d ocx @ 351 36 @ 3 @ 1

7.1.2 Insufficient vacuum

NOTE

The vacuum checks must be carried out when the ice condenser is frozen.

Pos: 168 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 169 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 20- 0010 Kleinfl anschv er bindu nge n @ 6\mo d_1 299 137 515 528 _68. docx @ 35 144 @ 4 @ 1

7.1.2.1 Small flange connections

Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
• Loosen the connection and place the centring ring (with sealing ring

inside) in a centred manner between the flange connections.

• Seal the connection with the clamping ring by tightening the wing nut.
• Ensure that the centring ring neither slips out of place nor gets jammed.

Fig. 65: Small flange and centring ring

Fig. 66: Small flange with centring ring
and small flange

Fig. 67: Attaching the clamping ring

Fig. 68: Tightened clamping ring