Semi-micro Calorimeter Instruction Manual

5-16

Recalculate Final Reports:
When set to ON, causes a recalculation of stored final reports using calibration data and
menu settings currently in the Calorimetric Thermometer. When set to OFF, the stored
final reports are not affected by the current run.

Use New EE Value in Recalculation:
When set to ON, any recalculation made will use the most recent EE value in the
calculations. In the OFF position, all calculations will be made using the EE value which
was effective when the test was originally run.

Communication Controls Menu

Accesses sub-menus which set the communications protocols for the printer and
balances.

PrinterType:

Toggles between a Parr 1758 and a generic printer. When set for the 1758
Printer, all of the features of this printer, such as bold printing, will be activated.


Balance Port:
Accesses sub-menu, Balance Port
Communications.

Balance Type:
Toggles through the available balance
templates.

Balance Port Device:
This key displays a screen which allows
the user to specify the balance port
device. The default (dev/ttyUSB0) is the
designation for the first USB to serial
converter cable assigned by the
calorimeter upon power up.

Customize Balance Settings:
Sets the communication parameters for the RS232 port used for the balance port.
Standard options for data bits, parity, stop bits, handshaking, baud rate and balance type
are provided to match any devices that might be connected to these ports.

• Number of Data Bits. Standard options for data bits. Toggles

between 7 and 8.

• Parity. Standard options for parity. Choose from None, Odd or

Even.

• Number of Stop Bits. Standard options for stop bits. Toggles

between 1 and 2.

• Handshaking. Standard options for handshaking. Choose from

Xon / Xoff, RTS/CTS and None

• Baud Rate. Standard options for baud rate. Choose from

19.2K , 9600, 4800, 2400, 1800, 1200, 600, 300, 150, 134.5,
110, and 75.