Default settings, Hapter – Parr Instrument 6755 User Manual

Page 50

Advertising
background image

14

Default Settings

P a r r I n s t r u m e n t C o m p a n y

48

Calorimeter Operation

Stirrer

OFF

Operating Controls

Method of Operation

Solution

LCD Backlight Timeout(s)

1200 S

LCD Backlight Intensity

70%

Print Error Messages

ON

Language

English

Program Info and Control

Settings Protect

OFF

Data Entry Controls

Auto Sample ID Controls

ON

Auto Sample ID Controls

Automatic Sample ID

ON

Automatic Sample ID Number

1

Automatic Sample ID Increment

1

Communication Controls

Printer

Printer Type

Parr 1758

Printer Destination

Local (USB)

Balance Port

Balance Port Device

/dev/tty/USB0

Number of Data Bits

8

Parity

None

Number of Stop Bits

1

Handshaking

None

Baud Rate

9600

Data Characters from Balance

8

Data Precision

4

Transfer Time Out

10

Diagnostics

Calorimetry Parameters

Tolerance (L) Parameters

Preperiod Tolerance (L1)

0.002

Misfi re Tolerance (L2)

0.05

Dynamic Post period Tolerance (L3)

0.002

Eqlbrm Post period Tolerance (L4)

0.002

Preperiod Stability (L5)

2

Preperiod Blackout (L6)

36

Derivative Wghtng Cutoff (L7)

0.02

Preperiod Timeout (L8)

42

Postperiod Timeout (L9)

60

Preperiod Tolerance Jck Calc’d (L10)

0.5

Correction (K) Parameters

K1

0.5

K2

0.00080

K3

1.0

K4

0.0

K5

0.0

Blackout (B) Parameters

Misfi re Blackout (B2)

72

Derivative Blackout (B3)

0.5

Dynamic Blackout (B4)

6

Equilibrium Blackout (B5)

18

Dynamic Derivative Blackout (B6)

0.02

Dynamic Time Blackout (B7)

20

Data Logger

Data Logger

OFF

Interval in Seconds

10

Data Log Destination

Log File and Printer

Data Log Format

Text

Reporting Controls

Report Width

40

Automatic Reporting

ON

Auto Report Destination

Printer

Individual Printed Reports

OFF

C

HAPTER

14

Default Settings

The following are recommended operation settings for the 6755 Solution Calorimeter.

Advertising