Identify the valves, Gas inlet valve, Gas release valve – Parr Instrument Series 4558 User Manual

Page 14: Liquid sampling valve

Advertising
background image

4555 - 58 Series

5G & 10L Pressure Reactors

- 13 -

IDENTIFY THE VALVES


Gas Inlet Valve

The gas inlet valve is easily identified when the vessel is open by noting that it is

connected to a dip tube which extends to a point near the bottom of the cylinder. Typically,
this is an angle valve with an attached fitting which provides a socket for attaching the
A495HC pressure hose furnished with the reactor.

Gas Release Valve

The gas release valve is connected to a plain opening on the underside of the head. Gas

released from this valve will be drawn from the top of the reactor.

Liquid Sampling Valve

The liquid sampling valve is attached to the same fitting as the gas inlet valve and

connected to the same dip tube. With this arrangement, incoming gas is always introduced
below the surface of the liquid and the operator is provided with a means for clearing the dip
tube to be sure that any sample taken during a run will be representative of the charge. This
can be done by opening the gas valve momentarily to force any liquid in the dip tube back
into the reactor before withdrawing a sample from the sampling valve.















Advertising
This manual is related to the following products: