Bottom Drain Valve Model A177VB2

P a r r I n s t r u m e n t C o m p a n y

2

Figure A-A

Assembly of Bottom Drain Valve

1. Before assembly, all parts of the valve should be

clean and in good condition.

Note:

Assembly of the valve is easier if it is removed

from service and oriented so that the threads going

into the cylinder face downwards.

2. Apply generous amounts of lubrication using

a high temperature thread lubricant. See the
“Lubrication of the Bottom Drain Valve” section
in this manual.

3. Install a 511VB5FH silver or 511VB5FG gold faced

(depending on BDV material) seal 1 into the cav-
ity in the valve body. The plunger may be used
to help with the installation.

4. Install the sleeve, and orientate it so the opening

is positioned over the valve drain opening.

5. Install the Graphoil seal 2 (front) and next the

PTFE seal 2 (back).

6. Install the retainer and start the (3) 5/16 socket

head cap screws.

7. Install the plunger assembly.

8. Tighten the retainer into valve body to the fol-

lowing torque:

A177VB2

100 in/lbs

9. Install the (3) standoffs.

10. Install the (3) 5/16 flat countersunk socket head

screws with the plunger retracted away from
seal 1.

11. Turn the shaft screw clockwise until the plunger

contacts seal 1.

BACK SEAL (PTFE)

FRONT SEAL (GRAPHOIL)

SEAL 1

SLEEVE OPENING

SLEEVE

DISASSEMBLY STEP 7

Advertising