Parr Instrument Primary Temp User Manual

Page 4

Advertising
background image

- 4 -

PTM Installation in a 4838

Factory Default Settings for Primary Temperature Module

Keys command:

1. Press “SET” to select
2. Press return key move to next operation mode

3. Up/Down arrow keys to adjust value or select type


Press return key and release

Operation
Mode

Select type/value

Comment

r-S Run

Run/Stop

SP

0**

Decimal point position

AL1H 375

Upper

alarm

setting

AL1L

0

Lower alarm setting

LoC

OFF

Lock mode(lock all keys or only up/down arrow

able to use)

Out1

- (read-only)

Heater output %

Out2

- (read-only)

Cooling output %

**

If using a type-J or type-K thermocouple, SV = 0

If using an RTD, SV = 1


Press and hold down “SET” for 5-sec


Operation
Mode

Select
type/value

Comment

InPt J**

Input

type

tPUn

C

Temperature unit

tP-H

800**

Upper-limit range

tP-L

0

Lower-limit temperature range

CTRL

PID

Control mode (ON/OFF, MANUAL, PID and PID
PROG)

S-HC

H1C2

Output 1 is Heating and output 2 is Cooling

ALA1

6

Alarm operates when PV value is higher than AL1H

setting or PV value is lower than AL1L setting

SALA OFF

System

alarm

CoSH

ON

Communication write function that able to use set
point from software

C-Sl ASCII Format

type

C-no 1

Communication

address

bPS

9600

Communication baud rate

Len

7

Data length setting

PrtY Even Parity

bit

setting

StoP

1

Stop bit setting


**

If using a type-J thermocouple, InPt = J, tP-H = 800

If using a type-K thermocouple, InPt = K, tP-H = 800

If using an RTD, InPt = Pt, tP-H = 600