Mcm installation, Factory default settings – Parr Instrument Motor Control Module (MCM) User Manual

Page 7

Advertising
background image

MCM Installation

w w w . p a r r i n s t . c o m

7

Factory Default Settings

MCM Module

Keys command:

1.

Press “SET” to select

2.

Press return key move to next operation mode

3.

Up/Down arrow keys to adjust value or select type

Main Screen: SP = 0

Press return key and release

Operation Mode

Select type/value Comment

r-S

Run

Run/Stop

SP

0

Decimal point position

LoC

OFF

Lock mode (lock all keys or only up/down arrow able to use)

Out1

- (read-only)

Motor output %

Press and hold down “SET” for 5-sec

Operation Mode

Select type/value Comment

InPt

v5

Input type
(v5 = 0-5V, v10 = 0-10V, nA0 = 0-20mA, nA4 = 4-20mA, nV =0-50mV)

tP-H

2206

Upper-limit range

tP-L

0

Lower-limit temperature range

CTRL

PID

Control mode (ON/OFF, MANUAL, PID and PID PROG)

S-HC

Heat

Heat/Cool control

ALA1

0

Alarm mode

SALA

OFF

System alarm

CoSH

ON

Communication write function that able to use set point from software

C-S1

RTU

Format type

C-no

2

Communication address

bPS

9600

Communication baud rate

Len

8

Data length setting

PrtY

Even

Parity bit setting

StoP

1

Stop bit setting

Press “SET” and release

Operation Mode

Select type/value Comment

At

OFF

Auto Tuning ON/OFF

PID0

0

The 0th PID Parameter

SV

0

The 0th SV

P0

170

Proportional control

C0

1

Integral control

D0

0

Derivative control

CoF0

0

Integral value

HtPd

.5

Heat/Cool cycle control

tPoF

0

Inaccuracy adjustment