OPERATING INSTRUCTIONS

THERMALERT MONITORING THERMOMETER, TH-5

1.0

The Physitemp Clinical Monitoring Instruments, Thermalerts TH-5 and TH-8, are compact,

portable, battery operated digital readout thermometers for continuous clinical temperature
monitoring.

2.0 INITIAL INSTALLATION

2.1

Plug any Physitemp type T thermocouple probe into the blue polarized socket on the rear

of the instrument.

2.2

The TH-5 thermometer is designed
to mount as follows:
(a) Flat, on rubber feet provided.
(b) Bench mounting with optional

tilt stand (TTS-4).

3.0

PRECAUTIONS

3.1

As with any temperature monitoring system, use suitable accepted technique during eletro-

surgical procedures to minimize the possibility that the probe cable will serve as an alternate
path for radio frequency current and burn the patient.

- Properly locate an adequate electrosurgical dispersive electrode close to the active

surgical site.

- Do not drape the probe cable over grounded metallic surfaces or intertwine it with the

cables of the electrosurgical generator.

3.2

Never pull directly on the probe cable to disconnect.

3.3

Gas sterilization or autoclaving of the instrument is not recommended.

3.4

Do not immerse the instrument in liquids for cleaning. If necessary, clean the instrument

case with a cloth moistened with a mild non-staining germicidal detergent and warm water.