2900 Indicator User Manual

Resolution Selection

*

Enter the resolution SET mode by pressing the

SHIFT/SET button followed by the RES button. The

choices are displayed sequentially from highest to

lowest. The round off accommodates the range of

values in the table (right).

Round Off

Condition

English

Inch Mode

Metric

mm Mode

Round Off Enabled

000.00005 0000.001

Round Off Disabled

000.0001 0000.001

Round Off Enabled

000.0005

0000.01

Round Off Disabled

000.001

0000.01

After selecting the desired resolution,

press the SET/SHIFT button to exit the

resolution mode. When the mode is

entered, the resolution selection is

presented in the units previously set

(English or metric). Press IN/mm to

toggle units selection, as required.

Cat. No.

EDP

Description

Resolution

Range

2900-1

09980 Basic English/Metric

.00005"/.001mm

.5"/12mm

2900-2

09981 Basic English/Metric

.0001"/.001mm

.5"/12mm

2900-4

09983 Standard English/Metric

with Selectable Resolution

.0005"/.0001"/.00005"

.01mm /.001mm

.5"/12mm

2900-6

09985 Advanced English/Metric

with Selectable Resolution

.0005"/.0001"/.00005"

.01mm /.001mm

.5"/12mm

2900-1M

09986 Basic Metric

.001mm

12mm

2900-4M

09988 Standard Metric

with Selectable Resolution .01mm /.001mm

12mm

2900-6M

09990 Advanced Metric

with Selectable Resolution .01mm /.001mm

12mm

2900 Indicator Models, Feature Level, Resolution & Range

Press

Then

to enter Resolution Mode

Press

to cycle through the available resolutions

to pick the resolution and exit resolution mode

Press

5

starrett.com

* Available with Standard & Advanced Features