Test Products International 163 User Manual

Page 16

Advertising
background image

16

17

FUNCTION BLACK RED

MINIMUM

MAXIMUM

TEST LEAD

TEST LEAD

READING

READING

µA

COM

mAµA

0.1µA

3260µA

mA

COM

mAµA

0.01mA

326mA

10A

COM

A

0.01A

10.00A

Measurement Procedure:

1.

Disconnect power to circuit to be measured.

2.

Plug the black test lead into the COM input jack.

3.

Plug the red test lead into the mAµA or A input jack
depending on the value of current to be measured.

4.

Set the rotary switch to the µA, mA, or 10A function.

5.

Connect test leads in series to circuit to be measured.

6.

Reconnect power to the circuit to be measured.

7.

Read the current on the 163.

d. MEASURING AC AMPS

CAUTION!
Do not attempt to make a current measurement with the
test leads connected in parallel with the circuit to be
tested. Test leads must be connected in series with the
circuit.

WARNING!
Do not attempt to make a current measurement of cir-
cuits with more than 600V present. Instrument damage
and/or personal injury may result.

Instrument set-up:

FUNCTION BLACK RED

MINIMUM

MAXIMUM

Measurement Procedure:

1.

Disconnect power to the circuit to be measured.

2.

Plug the black test lead into the COM input jack.

3.

Plug the red test lead into the V

input jack.

4.

Set the rotary switch to the V function.

5.

Connect the test leads to the circuit to be measured.

6.

Reconnect power to the circuit to be measured.

7.

Read the voltage on the 163.

C. MEASURING DC AMPS

CAUTION!
Do not attempt to make a current measurement with the
test leads connected in parallel with the circuit to be
tested. Test leads must be connected in series with the
circuit.

WARN-
ING!
Do not
attempt to
make a
current measurement of

circuits with

more than 600V present. Instrument

damage and

/or personal injury may result.

Instrument set-up:

FUNCTION BLACK RED

MINIMUM

MAXIMUM

TEST LEAD

TEST LEAD

READING

READING

µA

COM

mAµA

0.1µA

3260µA

Application Notes

When measuring the

DC current of a flame

controller, follow the steps under
“Measurement Procedure” below and
then proceed with the following:

Set up the meter for making a
mA measurement.

Connect the meter to the flame
controller lead by opening the
circuit and inserting the leads in
series with the circuit as shown
in the picture above.

Advertising