b. Measuring AC Volts

WARNING!

Do not attempt to make a voltage measurement of
more than 600V or of a voltage level that is unknown.

Instrument set-up:

FUNC.

BLACK RED

MIN

MAX

TEST LEAD TEST LEAD

READING

READING

V

COM

V/

0.01V

600V

Measurement Procedure:

1.

Disconnect power to circuit to be measured.

2.

Plug black test lead into COM input jack.

3.

Plug red test lead into V/

Ω input jack.

4.

Set the rotary switch to the V function.

5.

Press the FUNC button until AC is on LCD.

6.

Connect test leads to circuit to be measured.

7.

Reconnect power to circuit to be measured.

8.

Read the voltage on the 293.

Optional Modes
TRIM:

Push TRIM button to stabilize reading.

HOLD:

Freezes the reading on the LCD.

RANGE:

Manually ranges the instrument.

PEAK:

Measures the peak voltage.

Application Notes

Disconnect power from the
terminal block, find the fuse or
circuit breaker that controls

the block and turn it off.

Set up the meter following the

steps under “Measurement

Procedure” on page 17. Then

proceed with the following:

• Connect the red test lead to

the hot side of the block and the

black lead to the neutral side of the block.
Reconnect power to the block and read the
voltage on the meter. The reading should be
approximately 110V to 130V.

Disconnect power from the block and move
the red wire to ground. Reconnect power to
the block and read the voltage on the meter.
Typically less than 20V should exist from
neutral to ground. If 110V or above exists,
the block may be wired incorrectly.

!