Appendix E: ERROR CODES & TROUBLESHOOTING

The 712 analyzer will display certain codes to let you know of a malfunction.

*Sensor replacement requires calibration gas.

Appendix D: INSTALLING THE OPTIONAL A773 OIL FILTER

When performing combustion tests on oil fired equipment it is important to use
the optional A773 oil filter to ensure stable readings. The A773 does not have to
be removed when working with other types of fuels.

PROCEDURE

1. Pull the yellow thermocouple out of the cord channel in the tubing up to the

cord stop.

2. Cut the tubing approximately 1 1/2 inches from the end of the A762 water trap / fil-

ter assembly.

3. Insert the A773 oil filter into the tube below the A762.
4. Cut the remaining piece of tube into a 1 1/2 inch piece and connect it onto the end

of the A773 oil filter.

5.. Press the thermocouple cord into the cord channel.

The A773 should be replaced when most of the pellets become discolored.

Code

Displayed

Code Definition

Possible Causes

Corrective Action

Flo
Err

Pump not drawing sample at
correct flow rate.

Blockage / kink in flue probe
hose.

Dirty or blocked filter(s).

Worn pump.

Check and rectify.

Replace filter(s).

Return to TPI for service.

InIt

O2

Err

Oxygen sensor failed to ini-
tialize

Flue probe connected to 712
prior to power up.

712 did not purge completely
from last sample.

Worn or defective oxygen
sensor.

Disconnect probe and
restart.

Purge for 20 minutes and
restart.

Return to TPI for sensor
replacement or install new
sensor.*

Try manually initializing the
sensor. See Appendix G.

InIt
CO
Err

Carbon monoxide sensor
failed to initialize.

Flue probe connected to 712
prior to power up.

712 did not purge completely
from last sample.

Worn or defective carbon
monoxide sensor.

Disconnect probe and
restart.

Purge for 20 minutes and
restart.

Return to TPI for sensor
replacement or install new
sensor.*

Try manually initializing the
sensor. See Appendix G.

Lo

bat

Low battery.

Battery needs to be charged. Charge battery. If the battery

won’t hold a charge, replace
the battery.

oFL

Overflow indication. The
pressure being measured is
outside the maximum mea-
surement capability.

Pressure being measured is
too high or low.

Pressure sensor damaged or
defective.

Remove pressure source.

Return to TPI for service.

oFL

Overflow indication. The tem-
perature being measured is
outside the maximum mea-
surement capability.

Temperature being measured
is too high or low.

Remove pressure source.

oPEn

Unable to read thermocouple
(temperature).

Temperature probe not con-
nected to input.
Worn temperature sensor.

Connect temperature probe
to analyzer.
Replace temperature probe
or flue probe.