Trouble Shooting Guide

Problem

Probable Cause

Possible Remedy

Instrument won’t
turn on

Flat batteries

Check and replace
batteries

‘SEnS Err’ dis-
played on Screen

Faulty Combustible
Gas Sensor

Combustible Gas
Sensor needs re-
plaing

CO error is dis-
played

Faulty CO sensor

Return Instrument
to supplier

3

General Specifications


Sensor Specifications

Dimensions

240mm X 74mm X

55mm

Goose Neck Length

16’’

Battery

1.5V Alkaline battery

C size X 2

Power Consumption

About 400mW in

typical use.

Sensor Life

CO Sensor

Months until 80

% original signal

(24 month

warranted)

Pressure Range

80 ~ 120kPa

Alarms Visual/Audible

Response Time

CO

90% of reading

in 30 seconds

Leak

Instantaneous

Operating Temperature

0

°C ~ 50°C

Type Resolution Range

Accuracy

Leak <50ppm

0~50,000ppm ±10%

CO 1ppm

0~2000ppm

±5ppm or 5%

whichever is

greater

14

Trouble Shooting Guide

Problem

Probable Cause

Possible Remedy

Instrument won’t
turn on

Flat batteries

Check and replace
batteries

‘SEnS Err’ dis-
played on Screen

Faulty Combustible
Gas Sensor

Combustible Gas
Sensor needs re-
plaing

CO error is dis-
played

Faulty CO sensor

Return Instrument
to supplier

3

General Specifications


Sensor Specifications

Dimensions

240mm X 74mm X

55mm

Goose Neck Length

16’’

Battery

1.5V Alkaline battery

C size X 2

Power Consumption

About 400mW in

typical use.

Sensor Life

CO Sensor

Months until 80

% original signal

(24 month

warranted)

Pressure Range

80 ~ 120kPa

Alarms Visual/Audible

Response Time

CO

90% of reading

in 30 seconds

Leak

Instantaneous

Operating Temperature

0

°C ~ 50°C

Type Resolution Range

Accuracy

Leak <50ppm

0~50,000ppm ±10%

CO 1ppm

0~2000ppm

±5ppm or 5%

whichever is

greater