General description, Operation – Test Products International A771 User Manual

Page 3

Advertising
background image

GENERAL

DESCRIPTION

The A771 is an advanced state-of-the-art Carbon
Monoxide monitor that may be used to measure
levels of carbon monoxide (CO) in th part per
million (ppm) range when used with a digital
multimeter (DMM).
The instrument has both audible and visual alarms
when CO has reached the preset level. This unit
may be operated as an alarm only personal
monitor.

NOTE: The product may be used with any DMM
capable of measuring 2-4 volts DC. The signal
output is 1 mV DC = 1 ppm CO.

The A771 uses and advanced electrochemical
sensor that measures Carbon Monoxide only. A
built-in filter reduces most cross-sensitivity to
other gases. This product should be used with
DMM’s with greater than or equal to 10 megohm
impedance.

OPERATION

(Investigation Use)

1. Attach the banana jack to the COM and V input jacks

of the DMM.

2. Turn on the DMM and select the 2 or 4 volt range

DC.

3. If the voltage reading is below 4 VDC with the A771

in the off position, replace the 9 volt battery in the
A771.

4. Turn on the A771. The alarm LED with flash and the

green LED should stay on.

5. In a clean environment, adjust the zero knob until

the DMM reads zero.

6. Place the A771 in the area to be tested and read the

CO in ppm on the DMM.

7. To deactivate the alarm sound, set the slide switch

to MUTE.

OPERATION

(Monitor Use)

1. Follow steps 1-5 above.
2. Unplug the A771 from the DMM.
3. Place the A771 in the belt pouch with the sensor

facing outward.

4. If an alarm occurs, use a DMM to measure the

concentration.

5. The red LED will illuminate continuously during

alarms