Casella CEL CEL-283 User Manual

Page 6

Advertising
background image

Casella USA
17 Old Nashua Road #15
Amherst, NH 03031-2839

www.casellausa.com

Page 6 of 6

tel:

(

800) 366-2966

[email protected]

23 Dec 2005

fax: (603) 672-8053

Specification

Accuracy
Sound level meter: IEC 651, BS 5969 and ANSI
S1.4 in the type 2 category. DIN 45 634
Maximum RMS hold and energy average as
Leq, Lim or LTm (3 or 6 second), 40dB dynamic
range linear scaled meter.
Ranges
30 - 70 dB, 60 - 100 dB and 90 - 130 dB
Overload indicator provided.
Measuring Limits
Self-noise less than 25dB(A) or less than
40dB(Flat). 135dB absolute maximum.
Microphone
17 mm (0.67 inch) electret condenser type
Frequency Response
dB(A) or Flat
Display
47 mm (1.9 inch) meter movement covering 40
dB dynamic range.
Time Weightings
Fast, Slow and Impulse as per standards.
Maximum Hold
Analogue hold of maximum RMS level with a
decay rate of less than 1 dB/5 min at 20ºC.
LEQ Calculation

By means of 10msec samples of RMS level.
Minimum measurement duration 1 second,
maximum greater than 24 hours.
Calibration
Field calibration checks by means of CEL-184 or
CEL-182 Acoustic Calibrators (NB coupler type
CEL-3379 necessary for X version).
Auxiliary Output
3.5 mm jack feed of conditioned AC output at
1.0V for FSD via 3k3 Ohms.
Batteries
3 x 6 F22 (or equivalent) - 2 for sound level
meter and one for calibrator.
Temperature Range
-10 to +50ºC operational. -15 to +60ºC storage.
Humidity Range
30% to 90% for ±0.5 dB
Electromagnetic Interference
<MSD for 400 A/M
Vibration Interference
<62d B (Flat) for 1 m/sec/sec
Dimensions
235mm x 75mm x 54mm (9.2in x 3in x 2.1in)
Weight
425g (15oz)

REFERENCE CONDITIONS FOR
CALIBRATION

Type of sound field
Free

Reference incidence direction
Microphone diaphragm

Reference pressure level
94 dB relative to 0.00002 Pa

Reference frequency
1 kHz
Reference measurement range
60-100

Reference temperature
20ºC

Date ........................

Calibration Certificate

CEL-283 serial number ...................... .... ... .

CEL-184 serial number ……………………….

Date …………………………………………….

CEL Instruments are manufactured and serviced
by:

Casella CEL Inc.

17 Old Nashuia Road #15,

Amherst, NH 03031 USA

PHONE (800) 366-2966

FAX (603) 672-8053

[email protected]



In line with our policy of continual design
improvement we reserve the right to amend any
of the details in this handbook.

Advertising