Casella CEL CEL-490 User Manual

Print on a3

Advertising
background image

Preparation

DO NOT attempt to power from an external source

without first consulting Section 1.8 of the handbook.

Install 4 x AA size batteries in the compartment in the

underside of the unit.
Screw the Class 1 microphone “finger-tight” on to the

preamplifier.
Connect the Class 1 preamplifier plus microphone or the

Class 2 microphone/preamplifier assembly by inserting it

into the socket in the top of the instrument case with

the red dot facing the front to ensure correct pin location.

Menus & Control

All operations are based upon a series of menus

arranged as shown in the figure.
A combination of function and navigator keys on the

front panel are used to move around the menu system

and implement commands. The functions are as follows.

Power ON/OFF.

and

Navigator keys that move the highlight around the

options.

and

Navigator keys that change the contents of a field.
Enable/Disable Quick Edit.
Display Main menu (may require several presses).
Accesses data screens.
On screen show active keys.
Shows that Quick Edit mode is available.

Menu options are implemented in two ways.
1.

Leave the required option highlighted.

2.

Use the quick edit function to change the option.
(Press

to activate and deactivate the function).

Configuring the Instrument

The instrument configuration specifies how the

instrument communicates with the operator and via the

RS232 interface.

Press

to switch the instrument ON and obtain the

Calibration Check screen.
Press

to see the Main menu which gives access to

further menus that select Measurement Mode, Setup,

Calibration, Data Recall from Memory and Configuration.
An indication such as 2/6 at the bottom of a screen

shows that the second of six possible options is

highlighted.
Press

to move to Configure.

Press

to see the configuration menu.

Make choices about the RS232 Communication Baud

rate, Menu Language, Backlight operation and

Microphone Response (Free Field or Random) as

appropriate by leaving the required choice highlighted on

the screen.
The instrument Time and Date, Graph Range and AC

Output for recording can also be set from the

Configuration menu.
Press

(several times) to return to the Main menu.

Calibration

Calibration ensures that the instrument is measuring

noise accurately. The Calibration menu is displayed at

the end of the self test sequence.

If “Scale” is mentioned, move to the Main menu, select
Calibration, then press

to see the Calibration menu.

Insert the 1/2" microphone far enough into the calibrator

cavity to be in contact with the stop in its’ cavity.
Insert the 1/4" microphone far enough into the coupler

cavity to be in contact with the stop in its’ cavity and the

microphone plus coupler far enough into the calibrator

cavity to be in contact with the stop in its’ cavity.

Support the calibrator and sound level meter so that the

microphone points vertically upwards to avoid distorting

the calibration volume.

Switch the calibrator ON.

When an incorrect level is shown, press

to set the

display to indicate 114.0 dB automatically and YES to

save the calibration.
If the microphone and response set (Free Field/Random)

require that some other value should be set, perform a
manual calibration. Use

and

to select Auto, press

to select Last Cal, now use

and

to adjust the

display to the correct value. Press

to save the

calibration.

Select Mode

The Measure Mode specifies the Bandwidth to be used

for measurement.
In the Main menu, move to Measure Mode and press

to see the Measure Mode menu.

Select a bandwidth for use by highlighting it.

Alternatively, use the Setup option (below) to select a

pre-stored setup for use.
Press

to return to the Main menu.

Select Setup

The Setup specifies the particular parameters that are to

be measured.
In the Main menu, move to Setup and press

to see

the Setup menu.
Select a pre-saved setup for use by highlighting it.
When a setup is to be prepared and saved, press

to

move to a setup options menu. This allows the following

measurement parameters to be specified.

Broadband parameters:
SLM Response:

RMS-, time- and peak-weightings, plus

energy exchange rate Q.

Period Results:

Interval between measurements,

Profiles for up to 15 broadband parameters.

The maximum Run Time for the selected

combination will be shown

Measured Functions:

Up to 15 Broadband Period parameters,

Up to 14 Broadband Cumulative parameters.

Dose Results:

Threshold, for including values in calculation

Normalisation period, to allow results from

different durations to be compared.

Narrowband parameters (when available):
SLM Response:

RMS-, time- and peak-weightings, plus

energy exchange rate Q.

Period Results (CEL-490 only):

Interval between measurements,

Note that no profiles are available for narrowband

measurement.

Measured Functions:

Up to 10 Narrowband Period parameters,

Up to 9 Narrowband Cumulative parameters.

Set Timers

From the Main menu, select Timers, to see a Timers

menu that gives the option of switching the Timers Off,

setting a Duration Timer, or setting a Delay only on a

CEL-490.
Make choices as appropriate by leaving them high-

lighted on the screen.

Broadband Measurement

When broadband measurement is selected, press

several times from any Menu screen until a broadband

measurement screen is displayed. Broadband screens

show one principal and four secondary parameters that

can all be changed by means of the Quick Edit function.
Press

to start a measurement run and store results.

If a clock is shown, a timed measurement has been set,

refer to Section 3.1 of the Handbook.

Only data stored during the current run can be reviewed
while more data is being collected. Press

then use

the navigator keys to review data screens.
When sufficient data has been inspected, press

to

return to the measurement screen.
When the run is to be stopped, press

and see the

following options:

Stop run (and store data),

Restart run (abandon the current run and start a

new run),

Continue run (continue the interrupted run).

Select an option and confirm it by pressing

.

Narrowband Measurement

When narrowband measurement is selected, press

several times from any Menu screen until a narrowband

measurement screen is displayed.
Narrowband screens show a single identified parameter

with a 140 dB measurement range.
The RMS weighting for the first broadband column is

the weighting used for all narrowband columns.

Preparation

DO NOT attempt to power from an external source

without first consulting Section 1.8 of the handbook.

Install 4 x AA size batteries in the compartment in the

underside of the unit.
Screw the Class 1 microphone “finger-tight” on to the

preamplifier.
Connect the Class 1 preamplifier plus microphone or the

Class 2 microphone/preamplifier assembly by inserting it

into the socket in the top of the instrument case with

the red dot facing the front to ensure correct pin location.

Menus & Control

All operations are based upon a series of menus

arranged as shown in the figure.
A combination of function and navigator keys on the

front panel are used to move around the menu system

and implement commands. The functions are as follows.

Power ON/OFF.

and

Navigator keys that move the highlight around the

options.

and

Navigator keys that change the contents of a field.
Enable/Disable Quick Edit.
Display Main menu (may require several presses).
Accesses data screens.
On screen show active keys.
Shows that Quick Edit mode is available.

Menu options are implemented in two ways.
1.

Leave the required option highlighted.

2.

Use the quick edit function to change the option.
(Press

to activate and deactivate the function).

Configuring the Instrument

The instrument configuration specifies how the

instrument communicates with the operator and via the

RS232 interface.

Press

to switch the instrument ON and obtain the

Calibration Check screen.
Press

to see the Main menu which gives access to

further menus that select Measurement Mode, Setup,

Calibration, Data Recall from Memory and Configuration.
An indication such as 2/6 at the bottom of a screen

shows that the second of six possible options is

highlighted.
Press

to move to Configure.

Press

to see the configuration menu.

Make choices about the RS232 Communication Baud

rate, Menu Language, Backlight operation and

Microphone Response (Free Field or Random) as

appropriate by leaving the required choice highlighted on

the screen.
The instrument Time and Date, Graph Range and AC

Output for recording can also be set from the

Configuration menu.
Press

(several times) to return to the Main menu.

Calibration

Calibration ensures that the instrument is measuring

noise accurately. The Calibration menu is displayed at

the end of the self test sequence.

If “Scale” is mentioned, move to the Main menu, select
Calibration, then press

to see the Calibration menu.

Insert the 1/2" microphone far enough into the calibrator

cavity to be in contact with the stop in its’ cavity.
Insert the 1/4" microphone far enough into the coupler

cavity to be in contact with the stop in its’ cavity and the

microphone plus coupler far enough into the calibrator

cavity to be in contact with the stop in its’ cavity.

Support the calibrator and sound level meter so that the

microphone points vertically upwards to avoid distorting

the calibration volume.

Switch the calibrator ON.

When an incorrect level is shown, press

to set the

display to indicate 114.0 dB automatically and YES to

save the calibration.
If the microphone and response set (Free Field/Random)

require that some other value should be set, perform a
manual calibration. Use

and

to select Auto, press

to select Last Cal, now use

and

to adjust the

display to the correct value. Press

to save the

calibration.

Select Mode

The Measure Mode specifies the Bandwidth to be used

for measurement.
In the Main menu, move to Measure Mode and press

to see the Measure Mode menu.

Select a bandwidth for use by highlighting it.

Alternatively, use the Setup option (below) to select a

pre-stored setup for use.
Press

to return to the Main menu.

Select Setup

The Setup specifies the particular parameters that are to

be measured.
In the Main menu, move to Setup and press

to see

the Setup menu.
Select a pre-saved setup for use by highlighting it.
When a setup is to be prepared and saved, press

to

move to a setup options menu. This allows the following

measurement parameters to be specified.

Broadband parameters:
SLM Response:

RMS-, time- and peak-weightings, plus

energy exchange rate Q.

Period Results:

Interval between measurements,

Profiles for up to 15 broadband parameters.

The maximum Run Time for the selected

combination will be shown

Measured Functions:

Up to 15 Broadband Period parameters,

Up to 14 Broadband Cumulative parameters.

Dose Results:

Threshold, for including values in calculation

Normalisation period, to allow results from

different durations to be compared.

Narrowband parameters (when available):
SLM Response:

RMS-, time- and peak-weightings, plus

energy exchange rate Q.

Period Results (CEL-490 only):

Interval between measurements,

Note that no profiles are available for narrowband

measurement.

Measured Functions:

Up to 10 Narrowband Period parameters,

Up to 9 Narrowband Cumulative parameters.

Set Timers

From the Main menu, select Timers, to see a Timers

menu that gives the option of switching the Timers Off,

setting a Duration Timer, or setting a Delay only on a

CEL-490.
Make choices as appropriate by leaving them high-

lighted on the screen.

Broadband Measurement

When broadband measurement is selected, press

several times from any Menu screen until a broadband

measurement screen is displayed. Broadband screens

show one principal and four secondary parameters that

can all be changed by means of the Quick Edit function.
Press

to start a measurement run and store results.

If a clock is shown, a timed measurement has been set,

refer to Section 3.1 of the Handbook.

Only data stored during the current run can be reviewed
while more data is being collected. Press

then use

the navigator keys to review data screens.
When sufficient data has been inspected, press

to

return to the measurement screen.
When the run is to be stopped, press

and see the

following options:

Stop run (and store data),

Restart run (abandon the current run and start a

new run),

Continue run (continue the interrupted run).

Select an option and confirm it by pressing

.

Narrowband Measurement

When narrowband measurement is selected, press

several times from any Menu screen until a narrowband

measurement screen is displayed.
Narrowband screens show a single identified parameter

with a 140 dB measurement range.
The RMS weighting for the first broadband column is

the weighting used for all narrowband columns.

dB

114.Ø dB

Auto

L

ZF

v04013

Ø9JanØ4 11:28

dB

114.Ø dB

Auto

L

ZF

v04013

Ø9JanØ4 11:28

AFmx

dB

AF

Aeq

AFmn

Zpk

L

L

L

L

L

v04012

0

140

AFmx

dB

AF

Aeq

AFmn

Zpk

L

L

L

L

L

v04012

0

140

Quick Edit is

available

Quick Edit is

available

Print on A3

Fold as Shown

Fold Line

Fold Line

TOP

Cut Here

v04003

Class

2, 1/4"

Microphone

&

Preamplifier

Pull the
knurled

sleeve

outwards

to

release

the

connector

ON/OFF

Navigator

keys

Sta

rt/

Pause

/

Resta

rt

Access

data

Change
menu

Ena

ble

Quick

edit/

Select

Back

light

Sto

p

(needs
to confirm)

Protective
grill ( )

Class

1, 1/2"

Microphone

Preamplifier

Red

dot

do

not

remove

FIELD

GUIDE

to

CEL-450/490

Sound

Level

Meters

Th

e C

EL-450

and

CEL-490

Sound

Level Meters consi

st

of the

instr

ument

unit

preamplifier

and

microphone

show

n.

Advertising
This manual is related to the following products: