Casella CEL CEL-480 User Manual

Print on a3

Advertising
background image

Preparation

DO NOT attempt to power from an external source

without first consulting Section 1.8 of the handbook.

Install 4 x AA size batteries in the compartment in the

underside of the unit.
Screw the Type 1 microphone “finger-tight” on to the

preamplifier.
Connect the Type 1 preamplifier plus microphone or the

Type 2 microphone/preamplifier assembly by inserting it

into the socket in the top of the instrument case with

the red dot facing the front to ensure correct pin location.

Menus & Control

All operations are based upon a series of menus

arranged as shown in the figure.
A combination of function keys and navigator keys on

the front panel are used to move around the menu

system and implement commands. The functions are as

follows.

Power on/off.

and

Move the highlight around the options.

and

Change the contents of a field.
Enable/Disable Quick Edit.
Display Main menu (may require several presses).
Accesses data screens.
On screen show active keys.
Shows that Quick Edit mode is available.

Menu options are implemented in two ways.
1.

Leave the required option highlighted.

2.

Use the quick edit function (press

to activate)

to change the option.

Configuring the Instrument

The instrument configuration specifies how the

instrument communicates with the operator and via the

RS232 interface.

MENU

DATA

Press

to switch the instrument on and obtain the

Calibration Check screen.
Press

twice to see the Main menu which gives

access to further menus that select Measurement

Mode, Setup, Calibration, Data Recall from Memory and

Configuration.
A message such as 2/6 at the bottom of a screen

indicates that the second of six possible options is

highlighted.
Press

to move to Configure.

Press

to see the configuration menu.

Make choices about the RS232 Communication Baud

rate, Menu Language, Backlight operation and

Microphone Response (Free Field or Random) as

appropriate by leaving the required choice highlighted on

the screen.
The instrument Time and Date can also be set from the

Configuration menu.
Press

(several times) to return to the Main menu.

Calibration

Calibration ensures that the instrument is measuring

noise accurately. The Calibration menu is displayed at

the end of the self test sequence.

If “Scale” is mentioned, move to the Main menu, select
Calibration, then press

to see the Calibration menu.

Insert the 1/2" microphone far enough into the calibrator

cavity to be in contact with the shoulders in the cavity.
Insert the 1/4" microphone far enough into the coupler

cavity to be in contact with the shoulders in the cavity

and the microphone plus coupler far enough into the

calibrator cavity also to be in contact with the shoulders.

Avoid distorting the calibration volume by supporting the

calibrator and sound level meter on a horizontal surface.

Switch the calibrator on.

MENU

Depending on the microphone type and the response

set, the display should show one of these levels.

Microphone

Response

Calibration Level

1/2"

Free Field

114.0 dB

1/2"

Random

113.8 dB

1/4"

Free Field & Random

113.6 dB

If it does not, press

then use

and

to adjust

the display to the correct value.
Press

to save the calibration.

Select Mode

The measurement mode specifies the bandwidth to be

used for measurement. With a CEL-480, the mode also

specifies how measurements will be timed.
Either select a measurement mode, or use a pre-stored

setup (see below).

Instruments with only broadband measurement have no

Mode option.

In the Main menu, move to Mode (if available) and press

to see the Mode menu.

Select a bandwidth for use by highlighting it.
On a CEL-480 press

again to see a Timers menu for

the bandwidth that gives the option of switching the

Timers Off, setting a Duration Timer or setting a Delay

Timer.
Make choices as appropriate by leaving each one

highlighted on the screen.
Press

(several times) to return to the Main menu.

Select Setup

The Setup specifies the particular parameters that are to

be measured.
In the Main menu, move to Setup and press

to see

the Setup menu.
Select a setup for use by highlighting it.
When a setup is to be prepared and saved, press

to

move to a setup options menu.
This allows the following measurement parameters to

be specified (when available).
Mode: Broadband,

Octaves,

Third-octaves.

The following broadband parameters can be specified

(when available).
SLM Response:

RMS-, time- and peak-weightings, plus

energy exchange rate Q.

Measured Functions:

Up to 14 parameters.

Dose Results:

Threshold, for including values in calculation

Normalisation period, allows results from

different durations to be compared.

Period Results (CEL-480 only):

Interval between measurements,

Intervals for up to 2 profiles.

The following narrowband parameters can be specified

(when available).
Measured Functions:

Up to 3 functions.

Frequency Scanning Mode:

Manual, the user judges when the level is stable,

Auto, the SLM determines when the level is

stable,

Timed, measurements are taken at user pre-set

intervals,

Repeat, on the CEL-480 auto and timed scans

can be repeated automatically at user pre-set

intervals.

Broadband Measurement

From any menu press

several times until a broad-

band measurement screen is displayed.

Broadband measurement screens show one principal

and four secondary parameters that can all be changed

by the quick edit function.
Press

and

to change the range.

Press

to start a measurement run and store results.

If a clock is shown, a timed measurement has been set,

refer to Section 3.1 of the Handbook.

Data stored during the current (or other) runs can be

reviewed while more data is being collected.
Press

then use the navigator keys to review data

screens.
When sufficient data has been inspected, press

to

return to the measurement screen.
When the run is to be stopped, press

.

Three options are offered:

Stop run (and store data),

Restart run (abandon the current run and start a

new run),

Continue run (continue the interrupted run).

Select an option and confirm it by pressing

.

DATA

MENU

Preparation

DO NOT attempt to power from an external source

without first consulting Section 1.8 of the handbook.

Install 4 x AA size batteries in the compartment in the

underside of the unit.
Screw the Type 1 microphone “finger-tight” on to the

preamplifier.
Connect the Type 1 preamplifier plus microphone or the

Type 2 microphone/preamplifier assembly by inserting it

into the socket in the top of the instrument case with

the red dot facing the front to ensure correct pin location.

Menus & Control

All operations are based upon a series of menus

arranged as shown in the figure.
A combination of function keys and navigator keys on

the front panel are used to move around the menu

system and implement commands. The functions are as

follows.

Power on/off.

and

Move the highlight around the options.

and

Change the contents of a field.
Enable/Disable Quick Edit.
Display Main menu (may require several presses).
Accesses data screens.
On screen show active keys.
Shows that Quick Edit mode is available.

Menu options are implemented in two ways.
1.

Leave the required option highlighted.

2.

Use the quick edit function (press

to activate)

to change the option.

Configuring the Instrument

The instrument configuration specifies how the

instrument communicates with the operator and via the

RS232 interface.

MENU

DATA

Press

to switch the instrument on and obtain the

Calibration Check screen.
Press

twice to see the Main menu which gives

access to further menus that select Measurement

Mode, Setup, Calibration, Data Recall from Memory and

Configuration.
A message such as 2/6 at the bottom of a screen

indicates that the second of six possible options is

highlighted.
Press

to move to Configure.

Press

to see the configuration menu.

Make choices about the RS232 Communication Baud

rate, Menu Language, Backlight operation and

Microphone Response (Free Field or Random) as

appropriate by leaving the required choice highlighted on

the screen.
The instrument Time and Date can also be set from the

Configuration menu.
Press

(several times) to return to the Main menu.

Calibration

Calibration ensures that the instrument is measuring

noise accurately. The Calibration menu is displayed at

the end of the self test sequence.

If “Scale” is mentioned, move to the Main menu, select
Calibration, then press

to see the Calibration menu.

Insert the 1/2" microphone far enough into the calibrator

cavity to be in contact with the shoulders in the cavity.
Insert the 1/4" microphone far enough into the coupler

cavity to be in contact with the shoulders in the cavity

and the microphone plus coupler far enough into the

calibrator cavity also to be in contact with the shoulders.

Avoid distorting the calibration volume by supporting the

calibrator and sound level meter on a horizontal surface.

Switch the calibrator on.

MENU

Depending on the microphone type and the response

set, the display should show one of these levels.

Microphone

Response

Calibration Level

1/2"

Free Field

114.0 dB

1/2"

Random

113.8 dB

1/4"

Free Field & Random

113.6 dB

If it does not, press

then use

and

to adjust

the display to the correct value.
Press

to save the calibration.

Select Mode

The measurement mode specifies the bandwidth to be

used for measurement. With a CEL-480, the mode also

specifies how measurements will be timed.
Either select a measurement mode, or use a pre-stored

setup (see below).

Instruments with only broadband measurement have no

Mode option.

In the Main menu, move to Mode (if available) and press

to see the Mode menu.

Select a bandwidth for use by highlighting it.
On a CEL-480 press

again to see a Timers menu for

the bandwidth that gives the option of switching the

Timers Off, setting a Duration Timer or setting a Delay

Timer.
Make choices as appropriate by leaving each one

highlighted on the screen.
Press

(several times) to return to the Main menu.

Select Setup

The Setup specifies the particular parameters that are to

be measured.
In the Main menu, move to Setup and press

to see

the Setup menu.
Select a setup for use by highlighting it.
When a setup is to be prepared and saved, press

to

move to a setup options menu.
This allows the following measurement parameters to

be specified (when available).
Mode: Broadband,

Octaves,

Third-octaves.

The following broadband parameters can be specified

(when available).
SLM Response:

RMS-, time- and peak-weightings, plus

energy exchange rate Q.

Measured Functions:

Up to 14 parameters.

Dose Results:

Threshold, for including values in calculation

Normalisation period, allows results from

different durations to be compared.

Period Results (CEL-480 only):

Interval between measurements,

Intervals for up to 2 profiles.

The following narrowband parameters can be specified

(when available).
Measured Functions:

Up to 3 functions.

Scanning Mode:

Manual, the user judges when the level is stable,

Auto, the SLM determines when the level is

stable,

Timed, measurements are taken at user pre-set

intervals,

Repeat, on the CEL-480 auto and timed scans

can be repeated automatically at user pre-set

intervals.

Broadband Measurement

From any menu press

several times until a broad-

band measurement screen is displayed.

Broadband measurement screens show one principal

and four secondary parameters that can all be changed

by the quick edit function.
Press

and

to change the range.

Press

to start a measurement run and store results.

If a clock is shown, a timed measurement has been set,

refer to Section 3.1 of the Handbook.

Data stored during the current (or other) runs can be

reviewed while more data is being collected.
Press

then use the navigator keys to review data

screens.
When sufficient data has been inspected, press

to

return to the measurement screen.
When the run is to be stopped, press

.

Three options are offered:

Stop run (and store data),

Restart run (abandon the current run and start a

new run),

Continue run (continue the interrupted run).

Select an option and confirm it by pressing

.

DATA

MENU

dB

Ø9Jan98 12.22

Last Cal.

L

ZF

01031

dB

Ø9Jan98 12.22

Last Cal.

L

ZF

01031

AFmx

30

100

dB

AF

Aeq

AFmn

Zpk

L

L

L

L

L

01010

AFmx

30

100

dB

AF

Aeq

AFmn

Zpk

L

L

L

L

L

01010

Quick Edit is

available

Quick Edit is

available

Cut Here

Cut Here

Cut Here

Cut Here

ME

NU

DATA

01009

Type

2,

1/4"

Microphone

&

Preamplifier

Pull

the

knurled

sleeve

outwards

to

release

the

connector

ON/OFF

Navigator

key

s

Start/

Pause/
Restart

Access

data

Change
menu

Enables
Quick

edit

Back

light

Stop
(needs
to

confirm)

Protective
grill

(

)

Type

1,

1/2"

Microphone

Preamplifier

Red

dot

do

not

remove

FIELD

GUIDE

TO

CEL

-440/

480

Sound

Level

Meters

Print on A3

Fold as Shown

Fold Line

Fold Line

TOP

Advertising