Note: Use the foam windscreen when measurements are made outdoors. Do not blow into the

microphone opening because incorrect noise dose readings will result.

Normal Mode (Daily Dose)

Test Time: Up to 16 hours

Power ON:

• Press and

release the power button. The red 3200% LED is stead y,

followed by

one LED sweep.

• The previously measured dose displays for about 10 seconds while the

red 3200% LED remains steady.

• If the previous dose was 3200%, the red 3200% LED flashes.
• When the red 3200% LED turns off, a slow flashing of the green 25%

LED begins, indicating the start of the measurement.

• The dosimeter continuously displays the current dose, indicated by a

flashing LED.

• After 16 hours, the accumulated dose is displayed for about 8 minutes.

• During the display period, the red 3200% LED is steady and the

flashing LED indicates the dose. In cases where the accumulated
dose is 3200%, the red 3200% LED flashes
.

• At the end of 8 minutes, the dosimeter automatically shuts off.

• The dose is stored in memor y and is displayed when the device is

powered on again.

Power OFF:

• To power off the dosimeter prior to its automa tic shutdown: Press and

hold

the power button until a pair of lights flashes (about 3 seconds) then

release

the button.

• Wait 5 seconds before powering ON again.

Interpreting Results — Normal Mode

Operating Instructions

(continued)

Quick Check (Estimated Hourly Dose)

Test Time: 2 minutes

Note: Use the foam windscreen when measurements are made outdoors. Do not blow into the

microphone opening because incorrect noise dose readings will result.

Power ON:

• Press and

hold the power button for 5 seconds . The red 3200% LED

is steady, followed by

several LED sweeps. Release power button.

• The previously measured dose is displayed for about 10 seconds while

the red 3200% LED remains stead y.

• When the red 3200% LED turns off and the green 25% LED flashes,

the two-minute Quick Check starts.

• After 2 minutes, the steady red 3200% LED indicates the test is over.

The flashing LED indicates the estimated dose per hour.

During the display period, the red 3200% LED remains steady, unless
the estimated dose is 3200%, then the red 3200% LED flashes.

The estimated dose per hour displays for 1 minute, then the

dosimeter shuts off.

The dose is stored in memor y and is displayed when the device is

powered on again.

Power OFF:

• To power off the dosimeter prior to its automa tic shutdown: Press and

hold the power button until a pair of lights flashes (about 3 seconds)
then

release the button.

• Wait 5 seconds before powering ON again.

Interpreting Results — Quick Check (2-minute test)

LED

Dose %

Risk Factor

Hearing Protection Needed

green (slow flash) < 25%

No risk of hearing loss

NO

green (fast flash)

25%

No risk of hearing loss

NO

green

50%

1/2 daily dose reached

NO

yellow

100%

Limit of permissible exposure

YES

red

200%

2x allowable daily dose

YES

red

400%

4x allowable daily dose

YES

red

1600%

16x allowable daily dose

YES

red

3200%

32x allowable daily dose

YES

Muffled hearing and
ringing in the ears is
likely after unprotected
exposure

If the LED color is:

The estimated dose

The estimated dose

per hour is:

per 8 hours is:

green (slow flash) < 25% < 200%
green (fast flash)

25%

200%

green

50%

400%

yellow

100%

800%

red

200%

1600%

red

400%

3200%

red

1600%

red

3200%