Scale Programming

31

7.2

Item Code Maintenance

Item code maintenance includes viewing item codes in memory and deleting item memory.

View Item Codes in Memory

The following procedure allows viewing all of the item codes stored in memory but will not change any of the
information (unit weight, tare weight, part number, quantity, part name, setpoints, and lot number) stored with
these codes.

1. Press the

MODE

key. The

WEIGHT

display shows

ProG

and the

QUANTITY

display shows

C XX

.

2. Press the

CODE

key. The

QUANTITY

display shows the first item code programmed and stored in

memory, example:

Id 123

.

3. Press the

+

key to view the next item in memory and continue pressing the

+

key until the DC-190 beeps.

SPEC 7,

bits 0,1

Setpoint Type

Setpoint Designation

00

Percent quantity

Setpoint 1: Quantity
The quantity value can be no greater than 999999.

Setpoint 2: Percent of Setpoint 1
The percentage value can be no greater than 999%. The value for Setpoint 2 is calculated
by multiplying the value for Setpoint 1 by the percentage value entered for Setpoint 2.

NOTES:

• If Setpoint 1 is equal to 999999, then Setpoint 2 must be set less than or equal to

100%.

• Setpoints 3–6 are not available when Setpoint 2 is a percent setpoint.

01

Percent weight

Setpoint 1: Weight
The quantity value can be no greater than 999999.

Setpoint 2: Percent of Setpoint 1
The percentage value can be no greater than 999%. The value for Setpoint 2 is calculated
by multiplying the value for Setpoint 1 by the percentage value entered for Setpoint 2.

NOTE:

• If Setpoint 1 is equal to 999999, then Setpoint 2 must be set less than or equal to

100%.

• Setpoints 3–6 are not available when Setpoint 2 is a percent setpoint.

10

Upper and lower

quantity limit

Setpoint 1:

Quantity

Setpoint 2:

Quantity

Setpoints 3–6: Quantity

11

Upper and lower

weight limit

Setpoint 1: Weight
Must be a weight value less than or equal to the capacity of the scale.

Setpoint 2: Weight
Must be a weight value less than or equal to the capacity of the scale, but must be less
than Setpoint 1 value.

Setpoints 3–6: Weight
Must each specify a weight value less than the value of the preceding setpoint.

NOTES:

All weight-based setpoint values must be entered with a decimal and all trailing zeros. All percentage-based
setpoint values must be entered to the nearest whole number (no fractional percentages).

The DC-190 can program up to six setpoints by repeating Steps 2 through 6, SPEC 18, bit 0 through 2 (Number of
Setpoints) determines the number of setpoints. The six setpoints are TTL output for quantity or weight, but not
percent quantity or percent weight. These values may be programmed 1 through 6 low to high or 1 through 6 high to
low.

Table 7-1. Setpoint Programming for Quantity and Weight