Table 2

For all ME EQuIPMEnT and ME sYsTEMs

Guidance and manufacturer’s declaration – electromagnetic immunity
The BC 44 is intended for use in the electromagnetic environment specified below.

The customer or the user of the BC 44 should assure that it is used in such an environment.
IMMunITY test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile.

If floors are covered with synthetic material, the

relative humidity should be at least 30 %.

Power frequency

(50/60 Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at

levels characteristic of a typical location in a

typical commercial or hospital environment.

nOTE: U

T

is the a.c. mains voltage prior to application of the test level.