Beurer BM 85 User Manual

Page 139

Advertising
background image

139

Table 2

For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacture’s declaration – electromagnetic immunity
The BM 85 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 85 should assure that it is

used in such an environment.
Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Electrostatic discharge

(ESD) IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile. If

floors are covered with synthetic material, the relative

humidity should be at least 30 %.

Electrical fast

transient/burst

IEC 61000-4-4

± 2 kV for power supply

lines

± 1 kV for input/output

lines

± 2 kV for power supply

lines

± 1 kV for input/output

lines

Mains power quality should be that of a typical com-

mercial or hospital environment.

Surge IEC 61000-4-5

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

Mains power quality should be that of a typical com-

mercial or hospital environment.

Voltage dips, short inter-

ruptions and voltage

variations on power sup-

ply input lines

IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 s

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 s

Mains power quality should be that of a typical com-

mercial or hospital environment. If the user of the BM 85

requires continued operation during power mains inter-

ruptions, it is recommended that the BM 85 be powered

from an uninterruptible power supply or a battery.

Power frequency

(50/60 Hz) magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at levels

characteristic of a typical location in a typical commer-

cial or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.