3.0 Calibration cont.

To calibrate the pH

Clean pH probe tip.
See Section 8.0 (the pH probe does not require

cleaning before the fi rst use).

pH 7.0 calibration

a) Turn pH meter on. Rinse pH probe tip in fresh

water, shake off excess water and place in a

pH 7.0 calibration solution. Wait for at least

one minute or longer if required for reading to

stablize to a constant value.

b) Long press the ‘calibrate’ button. When the

display shows CAL release button.

pH 7 indicator is displayed indicating

a successful pH 7 calibration.

The pH 4 indicator will now fl ash

indicating pH 4.0 or pH 10.0

calibration is now required.

c) If Err appears during the calibration process see

Section 11.0.

d) The combo meter must be calibrated to two

points. If after an hour the combo meter has not

been calibrated with a second calibration point

the calibration indicators disappear and the

combo meter reverts to an uncalibrated state.

Calibration is required.

pH 4.0 / 10.0 calibration
a) Rinse the pH probe tip in fresh water, shake

off excess water and place the pH probe tip in

either pH 4.0 or pH 10.0 calibration solution.

b) Wait for at least one minute for reading to

stabilize to a constant value.

c) Long press the ‘calibrate’ button. When the

display shows CAL release button.

pH 7 / pH 4 is displayed

or pH 7 / pH 10 is displayed

d) The pH meter is now calibrated and ready for use.
e) After 30 days, the calibration indicators

disappear to let you know calibration is required.

Store the pH probe
Add enough Bluelab pH Probe KCl Storage Solution

into the probe storage cap to fully submerge the pH

probe tip. Place storage cap on probe.

1

2

3

4

pH 4.0 or

pH 10.0

calibration

solution

pH 7.0

calibration

solution

clean the pH probe

rinse in tap water

Ensure probe tip is covered by

the KCl storage solution in cap

Successful pH 7

and pH 4 calibration