7.0 Battery

replacement

Batteries are replaced in the Bluelab Soil pH Meter when the low battery indicator

appears on screen. The low battery indicator remains on and the Bluelab Soil pH

Meter continues to operate until the batteries die or are replaced.

Open battery compartment by sliding the back cover down and

insert 2 x AAA batteries as shown on the battery holder. Slide

cover back on. NOTE: Alkaline batteries are recommended.

IMPORTANT: Check the batteries at least once every six

months for signs of deterioration, rusting or swelling.
If signs of deterioration are found, clean battery holder

contacts and replace batteries.

1

2

Battery
cover

8.0 Cleaning the Bluelab Soil pH Probe

To ensure accurate readings the soil pH probe tip needs to be rinsed in water after each

use and cleaned prior to calibration using the following instructions.
The storage cap must always be put back on after cleaning. Always ensure it contains

enough Bluelab pH Probe KCl Storage Solution to cover the probe tip.

Remove storage cap from soil pH probe.
Hold the top of the storage cap, twist the cap to

loosen then remove.

Rinse soil pH probe tip under fresh tap water.
Never use RO (Reverse Osmosis), Distilled or

De-ionized water.

Fill a small plastic container with clean tap water.
Add a small amount of Bluelab pH Probe Cleaner

or mild detergent (dishwashing liquid).

Gently stir the probe tip in the mixture.
Ensure that you do not ‘knock’ the soil pH probe

on the side of the container as this may cause

damage to the probe.
Rinse well under fresh running water to remove all

traces of the detergent mixture.

If the probe tip requires removal of heavy

contamination: Gently brush around the

glassware with a few drops of Bluelab pH Probe

Cleaner or mild detergent (dishwashing liquid) and

a soft toothbrush.

Rinse well under fresh running tap water to

remove all traces of the detergent mixture.

Calibrate soil pH probe after cleaning, see

section 3.0 After calibration, store soil pH probe

in the storage cap, ensuring there is enough KCl

Storage Solution to cover the probe tip.