3 connection to valves, 4 connection to remote equipment – Bronkhorst R Modules for Ex-proof applications User Manual

Page 14

Advertising
background image

BRONKHORST HIGH-TECH B.V.

3.3 Connection to valves

The valve connector is a female sub-miniature 9-pin D-connector. The pin designation is according to the

Bronkhorst High-Tech B.V. standard for control valves.

pinnumber description

1

2

3

4

5

6

7

8

9

not connected

not connected

not connected

0 V valve

+ V valve

not connected

not connected

not connected

ground

Only XB or XC type valves may be connected. Consult the label on the module for the correct valve model.

3.4 Connection to remote equipment

For modules 20, 21 and 22:

The female in/out (1) (sub-miniature 9-pin) D-connector has the following pin configuration:

pinnumber description

1

2

3

4

5

6

7

8

9

output signal

input signal

0 V (common)

not connected

relay contact MC

relay contact NO

relay contact NC

reset input

ground (shield)

For modules 23, 24 and 25:

The I/O connector is a male sub-miniature 9-pin D-connector. The pin designation is according to the

Bronkhorst High-Tech B.V. standard for analog instruments.

pinnumber description

1

2

3

4

5

6

7

8

9

not connected

output signal

setpoint signal

not connected

not connected

default: not connected / optional : -15Vdc supply

default: +24Vdc supply / optional : +15Vdc supply

0Vdc / common

ground

Power supply (default +24Vdc) must be connected to pin 7 (+) and pin 8 (0Vdc / common)

Analog input signals should be connected to pin 3 (+) and 0 Vdc/common.

Analog output signals are available at pin 2 (+) and 0 Vdc/common.

Signals are according to one of the Bronkhorst High-Tech B.V. standards. The model configuration contains a

code, describing the input/output signals.

Notes:

a. Max. load current output (0 – 20 mA / 4 – 20 mA sourcing)

: 375 Ohm

b. Min. load voltage output (0 – 5 Vdc / 0-10 Vdc)

: 2 kOhm

c. Input load resistance (voltage 0 – 5 Vdc / 0-10 Vdc)

: 1M

d. Input load resistance (current 0 – 20 mA / 4 – 20 mA)

: 250 Ohm

The power supply of the modules 23, 24 and 25 is factory set to default +24Vdc.

page 14

9.17.034

Advertising