3.4.4 Alternatives for starting zero procedure

Through FLOW-BUS, using an E-7000 readout/control module.

Through FLOW-BUS, via a RS232/FLOW-BUS converter using software on a PC or PLC.

Through RS232, using software on a PC or PLC

Through RS232, using a Bright compact local readout/control module

Through other fieldbus system (PROFIBUS-DP/DeviceNet/ModBus)

3.4.5 Zeroing with the Micro-switch

Before using the instrument zeroing is required.

Set process conditions

After warm-up, pressure up and fill system including the mini CORI-FLOW

according to the process conditions.

Stop flow

Make sure there is no flow through the instrument by closing the shut-off valves before and after the

instrument.

Press and hold, until GREEN LED is on, than release button

With no flow, use the push-button switch (#) on the left side of the instrument to start the zero

adjustment procedure.

Press the push-button (#) and hold it, after a short time the red LED will go ON and OFF then the

green LED will go ON. At that moment release the push-button (#).

Zeroing

The zeroing procedure will start at that moment and the green LED will blink fast. The zeroing

procedure waits for a stable signal and saves the zero. If the signal is not stable, zeroing will take long

and the nearest point to zero is accepted. The procedure will take approx. 25 sec.

Make sure that there is no flow through the instrument when performing the zeroing procedure.

Ready

When indication is showing 0% signal and the green indication LED is continuously on, then zero has

been performed well.

3.4.6 Zeroing through the digital communication

It is also possible to start the zero adjustment procedure through the FLOW-BUS, using an E-7000

readout/control unit or a software program on a PC, connected to a FLOW-BUS interface module.

The following parameters must be used for zeroing an instrument:

Initreset

[unsigned char, RW,0...255, DDEpar. = 7, Proces/par. = 0/10]

Cntrlmode

[unsigned char, RW,0...255, DDEpar. = 12, Proces/par. = 1/4]

CalMode

[unsigned char, RW,0...255, DDEpar. = 58, Proces/par. = 115/1]

Set process conditions

Warm-up, pressure up the system and fill the instrument according to the process conditions.

Stop flow

Make sure there is no flow through the instrument by closing the shut-off valves before and after the

instrument.

Send parameters

Send the following values to the parameters in this sequence.

Initreset

64

Cntrlmode

9

Calmode

255

Calmode

0

Calmode

9

Zeroing

The zeroing procedure will start at that moment and the green LED will blink fast. The zeroing

procedure waits for a stable signal and saves the zero. If the signal is not stable zeroing, will take long

and the nearest point to zero is accepted. The procedure will take approx. 25 sec.

Make sure there is no flow through the instrument when performing the zeroing procedure.

Ready

When indication is showing 0% signal and the green indication LED is burning

continuously again, then zeroing has been performed well. Also parameter control mode

goes back to its original value. As last send 0 to parameter initreset.