Legal-for-Trade

Certification

D3709 Rev B

EZ2400 User’s Manual

25

Sealed to Prevent Unauthorized Modifications

The scale indicator is sealed by a Category 1 Audit Trail. To access the audit

trail in the normal weighing mode, press [NET/GROSS] and then press [ON]

within three seconds. This will display the current parameter and calibration

event count values. Pressing any key will return the indicator to the weighing

mode. The audit trail system is always active and cannot be altered. Audit trails

are stored in flash memory and do not require batteries to maintain the audit

trail values.

Changing any of the following parameters will change the Calibration event

count value or parameter audit event count value. This will void your Legal-for-

Trade certification unless performed by properly trained personnel.

Setup

Calibration

Capacity

Display

Count

Display

Units

Weigh

Method

Motion

WMA1-1 through WMA2-3

LFT Option Setting

On Legal-For-Trade Units Only

Perform the following procedure on an empty and clean machine before the

unit is to be “site-certified”.

Install the LFT indicator on the machine and connect all load cells.

Zero/balance the scale with no load on the machine.

Hold and press .

Press and to return to normal weighing.