Obtain current set-up and calibration number – Digi-Star EZ400 User Manual

Page 21

Advertising
background image

Short Form Calibration

D3655-US Rev E

EZ400 User’s Manual

17

SHORT FORM CALIBRATION

The Short Form Setup & Calibration procedure allows you to change “SETUP”

and “CAL” numbers of indicator.

Do not attempt to calibrate scale if indicator is not reading stable weights.

Calibration procedure will not fix instability, inconsistencies, or flashing

"RANGE" messages.

Obtain Current Set-up and Calibration Number

Write down current SETUP and CAL

numbers of your EZ 400 indicator.

These numbers are displayed during

Self Test.

SETUP NUMBER

To run self test with indicator ON:
1. Press then to start Self

Test.

2. Press to “pause” the Self-Test

while numbers are displayed.

3. Press again to allow self-test

to complete normally.

SETUP # _______________

CAL # ________________

Following is a list of functions that are controlled by the “SETUP” number:

Weigh Method (

W MTHD

)

Gain

Display Units (

LB-KG

)

Scale

Capacity

Display Counts (

COUNT

)

1

3

2

1

Advertising