Get your calibration number, Calibration, Enter – Digi-Star EZ4600 User Manual

Page 41: And press . the calibration, Number will display. example

Advertising
background image

Re-Calibrating Your Scale

D3716-US REV C

EZ4600 User’s Manual

37

SETTING

[display]

D.A.N

NO.

OPTIONS [displayed]

BOLD=DEFAULT

DESCRIPTION

CALIBRATION

SETUP NUMBER

(SETUP)

871

Quick entry method

selects weigh method 1-

4lbs, 5-8 kg, gain 1-9,

display counts 1-9 and

capacity *1000

CALIBRATION

NUMBER

(CAL)

872

Weight displayed at

0.4mV/V for these load

cells

RE-CALIBRATING YOUR SCALE

To re-calibrate your scale and make it even more accurate, document at

least 3 to 6 loads of varying sizes and measure the actual weight of all loads

on a certified scale.

Reading Too High

If the Indicator is reading higher than the certified scale, then the calibration

number is high and should be decreased proportionally.

Reading Too Low

If the Indicator is reading lower than the certified scale, then the calibration

number is low and should be increased proportionally.

Get your Calibration

Number

1. Enter

872

and press . The

calibration (

CAL

) number will

display. Example

CAL = 24280.

1

2 4 2 8 0

Press help for
more information

1

TOTAL CERTIFIED WEIGHT

TOTAL INDICATOR WEIGHT

X CURRENT CAL NUMBER=NEW CAL NUMBER