B. keys – Electro-Chemical Devices (ECD) L20 User Manual

Page 6

Advertising
background image

B. Keys

On/Off

Press and hold this key for 5 seconds to power on and

shut off the meter. Once the unit is power up, press the

same key to turn on or off the backlight.

Mode-

Selects display mode. Pressing this key changes the

display sequentially to display pH-AUTO, mV-AUTO, pH

and mV. The calibration values will not be affected by

changing the display modes. In “pH calibration”, press

“Mode” key to exit the two point calibration mode.

Clear-

It is used to clear the unit when error signal appears. It

clears all calibration values stored in the internal

memory.

Under normal use the key will not be activated unless

pressed and held for 2 seconds to prevent accidental

erasing stored memory.

When the Clear key is pressed, all segments of the LCD

will be on. After about 2 seconds the unit will enter the

pH-AUTO mode. The “AUTO” and “CAL” will be on and

one of the buffer values will start to flash. This means

that the unit must be calibrated.

Up/Down-

Increases or decreases the displayed temperature value

when in manual mode, buttons have no effect when in

ATC mode.

Stand/Slope-

The Stand and Slope keys are used for dual or three

point pH calibration of the unit. Pressing and holding

the Stand key while turning on the power, will change

the buffer set to the other available buffer set.

Mea. / Eff.-

The key is used to bring the unit out of the AUTOLOCK

condition when operating in the pH-AUTOLOCK or mV-

AUTOLOCK mode. Press and hold this key for 5 seconds,

the LCD will display the efficiency of the electrode.

6