Worcester controls – Flowserve 93 Series PULSAIR III User Manual

Page 6

Advertising
background image

6

Series 93 PULSAIR III Digital Electronic Positioner Installation, Operation and Maintenance

19860-C

Flow Control Division

Worcester Controls

14. Press the FUNC button once and the display should read

‘Transm’ over ‘Transm low’.

15. Press the FUNC button once and display should read ‘Transm

low’ over ‘LO=4.0 mA’.

16. Press the UP or DOWN button as necessary to adjust the output

signal to read 4.0 mA, or as close to 4.0 as is possible.

17. Press the OK button to accept the setting. The display should

read ‘Transm low’ over ‘UPDATED’ for a brief moment.

18. Press ESC and display should read ‘Transm’ over ‘Transm low’.

19. Press DOWN once and display should read ‘transm’ over

‘Transm hi’.

20. Press FUNC button once and the display should read ‘Transm hi’

over ‘HI=20.0 mA’.

21. Press the UP or DOWN button as necessary to adjust the output

signal to read 20.0 mA, or as close to 20.0 as is possible.

22. Press the OK button to accept the setting. The display should

read ‘Transm hi’ over ‘UPDATED’ for a brief moment.

23. Press the ESC button for 2-3 seconds and the display will return

to the ‘FULL MENU’ over ‘MAN/AUTO’ screen.

24. Disconnect the calibration ammeter from terminal positions 11

and 12, and attach your readout leads to terminal position 11
(positive) and 12 (negative).

Limit (M2) and Proximity (R2) Switch Cam Adjustment
(for L93W Positioner only)

1. Connect an ohmmeter to terminal position 3 (switch 1 -

normally open) and terminal position 5 (switch 1 - common).

2. Input a 4.2 mA signal so that the positioner is in the clockwise

position.

3. Using a screwdriver or similar tool, use the notch on the lower

cam and rotate the cam counter-clockwise away from the
switch and then clockwise until switch 1 closes. The ohmmeter
indication changes from infinite to a low reading that indicates
continuity through the switch.

4. Input a 19.8 mA signal so that the positioner is in the counter-

clockwise position.

5. Connect the ohmmeter leads to terminal position 6 (switch 2 -

normally open), and terminal position 8 (switch 2 - common).

6. Using a screwdriver or similar tool, use the notch on the upper

cam and rotate the cam clockwise away from the switch, and

then counter-clockwise as necessary until switch 2 closes. The
ohmmeter indication changes from infinite to a low reading that
indicates continuity through the switch.

7. Tighten the 2 mounting screws on the cam assembly to lock the

cam positions in place.

8. Adjust the input signal to 15 mA, and return to 20 mA to verify

that the cam contacts or trips the switch, and the ohm reading
changes from infinite to a low reading.

9. Remove the ohmmeter leads from terminal positions 6 and 8,

and attach them to terminal positions 3 and 5.

10. Adjust the input signal to 4 mA and verify that the cam contacts

or trips the switch and the ohm reading changes from infinite to
a low reading.

11. If the ohm reading does not change, and continues to show an

infinite reading, loosen the cam assembly mounting screws and
adjust the cam as necessary.

12. Repeat the adjustment and set procedure until you can supply a

4 mA and a 20 mA signal and have the switches close.

Namur Sensor (P2) Cam Adjustment (for L93W Positioner only)

Refer to diagram below:

1. Connect a milliammeter positive lead to terminal position 3

(sensor 1 - negative), and a 24 VDC power supply positive lead
to terminal position 4 (sensor 1 - positive). Connect the
millammeter negative lead to the 24 VDC power supply negative
terminal.

2. Input a 4.2 mA signal so that the positioner is in the clockwise

position.

3. Using a screwdriver or similar tool, use the notch on the lower

cam and rotate the cam counter-clockwise away from the sensor

Advertising